Drug & Device Safety Atlas
Class II FDA device recall · Z-2049-2026

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.5

Why it was recalled

For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
LUX-Dx II Plus UDI-DI: 00802526620706 (US), 00802526623707 (EU/AU); SERVER SW LATITUDE DRAGON EU UDI-DI: 00802526613821; SERVER SW LATITUDE DRAGON AU UDI-DI: 00802526613814; Serial Number: 100085 100676 100909 100944 101811 102341 102405 102864 102956 102978 103018 105283 105697 106372 106429 106690 107386 107460 108132 108429 108433 108641 108697 108805 108861 108901 109663 109699 109836 110700 110737 110858 111048 111119 111269 111290 111362 111780 111987 112265 112549 113400 113495 113720 114072 114393 115310 115411 115835 115862 115890 115981 116066 116090 116197 117180 117533 117560 117820 117924 117960 119462 120931 121080 121203 121267 121483 122855 123305 123412 124378 124458 124557 124791 124987 125008 126249 127182 127441 127758 127821 128106 128506 129528 129763 129782 130474 131469 132487 132549 133195 133786 133989 134600 135362 136038 136746 136884 137047 137134 137319 137449 138039 138059 138334 138759 139080 139165 140319 141181 142273 143441 143887 144590 144777 145284 145336 145684 146709 147185 147618 148972 149028 153330 153628 155199 155735 155898 156145 156895 157872 158926 159162 160236 160471 160615 161307 161762 162801 162821 163556 163802 167866 169673 169701 171284 172766 175538

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide distribution - US Nationwide and the country of Belgium.

Recall record

Recalling firm
Boston Scientific Corporation
Product type
Devices
Recall number
Z-2049-2026
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Saint Paul, MN
Recall initiated
2026-03-30

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2049-2026 · Source: openFDA enforcement.