Class I FDA device recall · Z-2173-2025

Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.

Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS

Why it was recalled

Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the current infusion rate, the duration the pump has been running at this flow rate, and the magnitude of the rate change. The longer the duration the pump has been running at the current infusion rate and the larger the magnitude of the rate change, the larger the underinfusion that would be experienced. Additionally, Baxter has identified an increase in customer reports of over and underinfusion potentially due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Consistent with the instructions for use, customers should ensure that: 1) The door is fully open before loading the set. 2) The tubing is taut and loaded without slack in the pumping channel.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI/DI 05413765851797, All Serial Numbers

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide distribution - US Nationwide and the country of Canada.

Recall record

Recalling firm
Baxter Healthcare Corporation
Product type
Devices
Recall number
Z-2173-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Deerfield, IL
Recall initiated
2025-07-14

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2173-2025 · Source: openFDA enforcement.