Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.
Automated Impella Controller (AIC); Product Code: 0042-0000-US;
Why it was recalled
A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
Product Code: 0042-0000-US; UDI-DI: 00813502010022; Serial Numbers: IC2449, IC2451, IC2452, IC2453, IC2459, IC2460, IC1096, IC1142, IC1202, IC1203, IC1224, IC1231, IC1238, IC1284, IC1285, IC1390, IC1545, IC1577, IC1665, IC1708, IC1741, IC1852, IC1853, IC2029, IC2071, IC2154, IC2277, IC2278, IC2314, IC2315, IC2368, IC2369, IC2370, IC2442, IC2443, IC2444, IC2461, IC2471, IC2476, IC2477, IC2478, IC2479, IC2575, IC2989, IC3127, IC3128, IC3129, IC3130, IC3166, IC3167, IC3213, IC3223, IC3306, IC3307, IC3308, IC3309, IC3342, IC3398, IC3475, IC3656, IC3658, IC3663, IC3664, IC3665, IC3666, IC3674, IC3675, IC3719, IC3720, IC3776, IC3812, IC3887, IC3888, IC4045, IC4060, IC1301, IC1544, IC1572, IC2686, IC2687, IC2848, IC3174, IC3269, IC3272, IC3277, IC3278, IC3281, IC3282, IC3284, IC3294, IC3783;
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide - US Nationwide distribution in the states of CA, CT, FL, LA, MA, MI, MO, MS, ND, NY, OH, PA, TX, WY and the countries of Brazil, Canada, Colombia, Hong Kong, Italy, Kuwait, Saudi Arabia, UAE, United Kingdom;
Recall record
- Recalling firm
- Abiomed, Inc.
- Product type
- Devices
- Recall number
- Z-2211-2026
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Danvers, MA
- Recall initiated
- 2026-04-20
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-2211-2026 · Source: openFDA enforcement.