  "@context": {
    "schema": "https://schema.org/",
    "au": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native#",
    "@vocab": "https://schema.org/",
    "claims": {
      "@id": "au:claims",
      "@container": "@list"
    },
    "epistemicStatus": "au:epistemicStatus",
    "occurred": "au:occurred",
    "supported": "au:supported",
    "adjudged": "au:adjudged",
    "operative": "au:operative",
    "grounding": "au:grounding",
    "groundingType": "au:groundingType",
    "adjudicator": "au:adjudicator",
    "source": "au:source",
    "sourceUrl": "au:sourceUrl",
    "value": "au:value",
    "claimType": "au:claimType",
    "reviewStatus": "au:reviewStatus",
    "commercialInfluence": "au:commercialInfluence",
    "falsifier": "au:falsifier"
  },
  "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-2213-2026.jsonld",
  "@type": "schema:Dataset",
  "name": "OneLIF Interbody Cage (REF/Description):  010-107-1007/Small 7 Degree 10mm X 7mm,  010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique,  010-107-1209/Small 7 Degree 12mm X 9mm,  010-107-1411/Small 7 Degree 14mm X 11mm,  010-107-1613/Small 7 Degree 16mm X 13mm,  010-113-1006/Small, 13 Degree 10mm X 6mm,  010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique,  010-113-1208/Small 13 Degree 12mm X 8mm,  010-113-1410/Small 13 Degree 14mm X 10mm,  010-113-1612/Small 13 Degree 16mm X 12mm,  010-119-1407/19 Degree 14mm X 7mm,  010-119-1609/Small 19 Degree 16mm X 9mm,  010-119-1811/Small 19 Degree 18mm X 11mm,  010-125-1606/Small 25 Degree 16mm X 6mm,  010-125-1808/Small 25 Degree 18mm X 8mm,  010-207-1007/Medium 7 Degree 10mm X 7mm,  010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique,  010-207-1209/Medium 7 Degree 12mm X 9mm,  010-207-1411/Medium 7 Degree 14mm X 11mm,  010-207-1613/Medium 7 Degree 16mm X 13mm,  010-213-1206/Medium 13 Degree 12mm X 6mm,  010-213-1408/Medium 13 Degree  14mm X 8mm,  010-213-1610/Medium 13 Degree 16mm X 10mm,  010-213-1812/Medium 13 Degree 18mm X 12mm,  010-219-1406/Medium 19 Degree 14mm X 6mm,  010-219-1608/Medium 19 Degree 16mm X 8mm,  010-219-1810/Medium 19 Degree 18mm X 10mm,  010-225-1807/Medium 25 Degree 18mm X 7mm,  010-225-2009/Medium 25 Degree 20mm X 9mm,  010-307-1006/Large 7 Degree,10mm X 6mm,  010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique,  010-307-1208/ Large 7 Degree 12mm X 8mm,  010-307-1410/Large 7 Degree 14mm X 10mm,  010-307-1612/Large 7 Degree 16mm X 12mm,  010-313-1205/Large 13 Degree 12mm X 5mm,  010-313-1407/Large 13 Degree 14mm X 7mm,  010-313-1609/Large 13 Degree 16mm X 9mm,  010-313-1811/Large 13 Degree 18mm X 11mm,  010-313-2013/Large 13 Degree 20mm X 13mm,  010-319-1606/Large 19 Degree 16mm X 6mm,  010-319-1808/ Large 19 Degree 18mm X 8mm,  010-319-2010/ Large 19 Degree 20mm X 10mm,  010-325-2007/Large 25 Degree 20mm X 7mm,  010-325-2209/ Large 25 Degree 22mm X 9mm,    Implant Kits containing OneLIF Interbody Cages, Name/Kit Family:  Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm,  Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,",
  "description": "FDA recall Z-2213-2026 — Novapproach Spine, LLC is recalling OneLIF Interbody Cage (REF/Description):  010-107-1007/Small 7 Degree 10mm X 7mm,  010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique,  010-107-1209/Small 7 Degree 12mm X 9mm,  010-107-1411/Small 7 Degree 14mm X 11mm,  010-107-1613/Small 7 Degree 16mm X 13mm,  010-113-1006/Small, 13 Degree 10mm X 6mm,  010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique,  010-113-1208/Small 13 Degree 12mm X 8mm,  010-113-1410/Small 13 Degree 14mm X 10mm,  010-113-1612/Small 13 Degree 16mm X 12mm,  010-119-1407/19 Degree 14mm X 7mm,  010-119-1609/Small 19 Degree 16mm X 9mm,  010-119-1811/Small 19 Degree 18mm X 11mm,  010-125-1606/Small 25 Degree 16mm X 6mm,  010-125-1808/Small 25 Degree 18mm X 8mm,  010-207-1007/Medium 7 Degree 10mm X 7mm,  010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique,  010-207-1209/Medium 7 Degree 12mm X 9mm,  010-207-1411/Medium 7 Degree 14mm X 11mm,  010-207-1613/Medium 7 Degree 16mm X 13mm,  010-213-1206/Medium 13 Degree 12mm X 6mm,  010-213-1408/Medium 13 Degree  14mm X 8mm,  010-213-1610/Medium 13 Degree 16mm X 10mm,  010-213-1812/Medium 13 Degree 18mm X 12mm,  010-219-1406/Medium 19 Degree 14mm X 6mm,  010-219-1608/Medium 19 Degree 16mm X 8mm,  010-219-1810/Medium 19 Degree 18mm X 10mm,  010-225-1807/Medium 25 Degree 18mm X 7mm,  010-225-2009/Medium 25 Degree 20mm X 9mm,  010-307-1006/Large 7 Degree,10mm X 6mm,  010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique,  010-307-1208/ Large 7 Degree 12mm X 8mm,  010-307-1410/Large 7 Degree 14mm X 10mm,  010-307-1612/Large 7 Degree 16mm X 12mm,  010-313-1205/Large 13 Degree 12mm X 5mm,  010-313-1407/Large 13 Degree 14mm X 7mm,  010-313-1609/Large 13 Degree 16mm X 9mm,  010-313-1811/Large 13 Degree 18mm X 11mm,  010-313-2013/Large 13 Degree 20mm X 13mm,  010-319-1606/Large 19 Degree 16mm X 6mm,  010-319-1808/ Large 19 Degree 18mm X 8mm,  010-319-2010/ Large 19 Degree 20mm X 10mm,  010-325-2007/Large 25 Degree 20mm X 7mm,  010-325-2209/ Large 25 Degree 22mm X 9mm,    Implant Kits containing OneLIF Interbody Cages, Name/Kit Family:  Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm,  Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit, (Devices, Class II). The enforcement record is RECORDED; the classification is a DETERMINATION (FDA's regulatory verdict); the boundary is INTERPRETATION (a recall ≠ harm to a specific person). Source: openFDA. CC0.",
  "license": "https://creativecommons.org/publicdomain/zero/1.0/",
  "isAccessibleForFree": true,
  "isPartOf": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
  "rdfs:isDefinedBy": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native",
  "seeAlso": "https://agenticu.ryanjhunter.workers.dev/commons/the-research-object",
  "dateModified": "2026-06-24T07:02:03.626072+00:00",
  "isBasedOn": {
    "@type": "schema:Dataset",
    "name": "openFDA drug & device enforcement (recall) reports",
    "url": "https://open.fda.gov/apis/drug/enforcement/",
    "creator": {
      "@type": "schema:GovernmentOrganization",
      "name": "U.S. Food & Drug Administration"
    }
  },
  "sameAs": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-2213-2026&lang=eng",
  "about": {
    "@type": "schema:MedicalDevice",
    "name": "OneLIF Interbody Cage (REF/Description):  010-107-1007/Small 7 Degree 10mm X 7mm,  010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique,  010-107-1209/Small 7 Degree 12mm X 9mm,  010-107-1411/Small 7 Degree 14mm X 11mm,  010-107-1613/Small 7 Degree 16mm X 13mm,  010-113-1006/Small, 13 Degree 10mm X 6mm,  010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique,  010-113-1208/Small 13 Degree 12mm X 8mm,  010-113-1410/Small 13 Degree 14mm X 10mm,  010-113-1612/Small 13 Degree 16mm X 12mm,  010-119-1407/19 Degree 14mm X 7mm,  010-119-1609/Small 19 Degree 16mm X 9mm,  010-119-1811/Small 19 Degree 18mm X 11mm,  010-125-1606/Small 25 Degree 16mm X 6mm,  010-125-1808/Small 25 Degree 18mm X 8mm,  010-207-1007/Medium 7 Degree 10mm X 7mm,  010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique,  010-207-1209/Medium 7 Degree 12mm X 9mm,  010-207-1411/Medium 7 Degree 14mm X 11mm,  010-207-1613/Medium 7 Degree 16mm X 13mm,  010-213-1206/Medium 13 Degree 12mm X 6mm,  010-213-1408/Medium 13 Degree  14mm X 8mm,  010-213-1610/Medium 13 Degree 16mm X 10mm,  010-213-1812/Medium 13 Degree 18mm X 12mm,  010-219-1406/Medium 19 Degree 14mm X 6mm,  010-219-1608/Medium 19 Degree 16mm X 8mm,  010-219-1810/Medium 19 Degree 18mm X 10mm,  010-225-1807/Medium 25 Degree 18mm X 7mm,  010-225-2009/Medium 25 Degree 20mm X 9mm,  010-307-1006/Large 7 Degree,10mm X 6mm,  010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique,  010-307-1208/ Large 7 Degree 12mm X 8mm,  010-307-1410/Large 7 Degree 14mm X 10mm,  010-307-1612/Large 7 Degree 16mm X 12mm,  010-313-1205/Large 13 Degree 12mm X 5mm,  010-313-1407/Large 13 Degree 14mm X 7mm,  010-313-1609/Large 13 Degree 16mm X 9mm,  010-313-1811/Large 13 Degree 18mm X 11mm,  010-313-2013/Large 13 Degree 20mm X 13mm,  010-319-1606/Large 19 Degree 16mm X 6mm,  010-319-1808/ Large 19 Degree 18mm X 8mm,  010-319-2010/ Large 19 Degree 20mm X 10mm,  010-325-2007/Large 25 Degree 20mm X 7mm,  010-325-2209/ Large 25 Degree 22mm X 9mm,    Implant Kits containing OneLIF Interbody Cages, Name/Kit Family:  Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm,  Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,",
    "manufacturer": {
      "@type": "schema:Organization",
      "name": "Novapproach Spine, LLC"
    }
  },
  "claims": [
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-2213-2026/enforcement-record",
      "@type": "schema:Claim",
      "name": "FDA recall Z-2213-2026: Novapproach Spine, LLC is recalling OneLIF Interbody Cage (REF/Description):  010-107-1007/Small 7 Degree 10mm X 7mm,  010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique,  010-107-1209/Small 7 Degree 12mm X 9mm,  010-107-1411/Small 7 Degree 14mm X 11mm,  010-107-1613/Small 7 Degree 16mm X 13mm,  010-113-1006/Small, 13 Degree 10mm X 6mm,  010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique,  010-113-1208/Small 13 Degree 12mm X 8mm,  010-113-1410/Small 13 Degree 14mm X 10mm,  010-113-1612/Small 13 Degree 16mm X 12mm,  010-119-1407/19 Degree 14mm X 7mm,  010-119-1609/Small 19 Degree 16mm X 9mm,  010-119-1811/Small 19 Degree 18mm X 11mm,  010-125-1606/Small 25 Degree 16mm X 6mm,  010-125-1808/Small 25 Degree 18mm X 8mm,  010-207-1007/Medium 7 Degree 10mm X 7mm,  010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique,  010-207-1209/Medium 7 Degree 12mm X 9mm,  010-207-1411/Medium 7 Degree 14mm X 11mm,  010-207-1613/Medium 7 Degree 16mm X 13mm,  010-213-1206/Medium 13 Degree 12mm X 6mm,  010-213-1408/Medium 13 Degree  14mm X 8mm,  010-213-1610/Medium 13 Degree 16mm X 10mm,  010-213-1812/Medium 13 Degree 18mm X 12mm,  010-219-1406/Medium 19 Degree 14mm X 6mm,  010-219-1608/Medium 19 Degree 16mm X 8mm,  010-219-1810/Medium 19 Degree 18mm X 10mm,  010-225-1807/Medium 25 Degree 18mm X 7mm,  010-225-2009/Medium 25 Degree 20mm X 9mm,  010-307-1006/Large 7 Degree,10mm X 6mm,  010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique,  010-307-1208/ Large 7 Degree 12mm X 8mm,  010-307-1410/Large 7 Degree 14mm X 10mm,  010-307-1612/Large 7 Degree 16mm X 12mm,  010-313-1205/Large 13 Degree 12mm X 5mm,  010-313-1407/Large 13 Degree 14mm X 7mm,  010-313-1609/Large 13 Degree 16mm X 9mm,  010-313-1811/Large 13 Degree 18mm X 11mm,  010-313-2013/Large 13 Degree 20mm X 13mm,  010-319-1606/Large 19 Degree 16mm X 6mm,  010-319-1808/ Large 19 Degree 18mm X 8mm,  010-319-2010/ Large 19 Degree 20mm X 10mm,  010-325-2007/Large 25 Degree 20mm X 7mm,  010-325-2209/ Large 25 Degree 22mm X 9mm,    Implant Kits containing OneLIF Interbody Cages, Name/Kit Family:  Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm,  Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,",
      "text": "The FDA logged enforcement record Z-2213-2026: Novapproach Spine, LLC is voluntarily recalling OneLIF Interbody Cage (REF/Description):  010-107-1007/Small 7 Degree 10mm X 7mm,  010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique,  010-107-1209/Small 7 Degree 12mm X 9mm,  010-107-1411/Small 7 Degree 14mm X 11mm,  010-107-1613/Small 7 Degree 16mm X 13mm,  010-113-1006/Small, 13 Degree 10mm X 6mm,  010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique,  010-113-1208/Small 13 Degree 12mm X 8mm,  010-113-1410/Small 13 Degree 14mm X 10mm,  010-113-1612/Small 13 Degree 16mm X 12mm,  010-119-1407/19 Degree 14mm X 7mm,  010-119-1609/Small 19 Degree 16mm X 9mm,  010-119-1811/Small 19 Degree 18mm X 11mm,  010-125-1606/Small 25 Degree 16mm X 6mm,  010-125-1808/Small 25 Degree 18mm X 8mm,  010-207-1007/Medium 7 Degree 10mm X 7mm,  010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique,  010-207-1209/Medium 7 Degree 12mm X 9mm,  010-207-1411/Medium 7 Degree 14mm X 11mm,  010-207-1613/Medium 7 Degree 16mm X 13mm,  010-213-1206/Medium 13 Degree 12mm X 6mm,  010-213-1408/Medium 13 Degree  14mm X 8mm,  010-213-1610/Medium 13 Degree 16mm X 10mm,  010-213-1812/Medium 13 Degree 18mm X 12mm,  010-219-1406/Medium 19 Degree 14mm X 6mm,  010-219-1608/Medium 19 Degree 16mm X 8mm,  010-219-1810/Medium 19 Degree 18mm X 10mm,  010-225-1807/Medium 25 Degree 18mm X 7mm,  010-225-2009/Medium 25 Degree 20mm X 9mm,  010-307-1006/Large 7 Degree,10mm X 6mm,  010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique,  010-307-1208/ Large 7 Degree 12mm X 8mm,  010-307-1410/Large 7 Degree 14mm X 10mm,  010-307-1612/Large 7 Degree 16mm X 12mm,  010-313-1205/Large 13 Degree 12mm X 5mm,  010-313-1407/Large 13 Degree 14mm X 7mm,  010-313-1609/Large 13 Degree 16mm X 9mm,  010-313-1811/Large 13 Degree 18mm X 11mm,  010-313-2013/Large 13 Degree 20mm X 13mm,  010-319-1606/Large 19 Degree 16mm X 6mm,  010-319-1808/ Large 19 Degree 18mm X 8mm,  010-319-2010/ Large 19 Degree 20mm X 10mm,  010-325-2007/Large 25 Degree 20mm X 7mm,  010-325-2209/ Large 25 Degree 22mm X 9mm,    Implant Kits containing OneLIF Interbody Cages, Name/Kit Family:  Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm,  Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm,  Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit,  Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit, — reason: Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.. Status: Ongoing — the recall is still being carried out.. Initiated 2026-04-23. Affected lots/codes: REF/UDI-DI/ Cage Lots: 010-107-1007/G07001010710070/ 103003, 105701, 118001, 129301; 010-107-1007-2/G070010107100720/ 114301, 117509, 128703; 010-107-1209/G07001010712090/ 102714, 105702, 117501; 010-107-1411/G07001010714110/ 103019, 105703; 010-107-1613/G07001010716130/ 103020, 105704, 127701; 010-113-1006/G07001011310060/ 103021, 105705, 118002, 129801; 010-113-1006-2/G070010113100620/ 117510, 127716, 114302; 010-113-1208/G07001011312080/ 103022, 105706, 117502, 126902; 010-113-1410/G07001011314100/ 102715, 105707, 127702; 010-113-1612/G07001011316120/ 103023, 105708, 127703; 010-119-1407/G07001011914070/ 103024, 105709, 125101, 128701; 010-119-1609/G07001011916090/ 102716, 105710, 127001; 010-119-1811/G07001011918110/ 103025, 105711, 127003; 010-125-1606/G07001012516060/ 102717, 117901, 127002; 010-125-1808/G07001012518080/ 103005, 117902, 127704; 010-207-1007/G07001020710070/ 102702, 106201, 118003, 125102; 010-207-1007-2/G070010207100720/ 117511, 124106, 127717, 114303; 010-207-1209/G07001020712090/ 102701, 105401, 124101, 127705, 101201; 010-207-1411/G07001020714110/ 105001, 105402, 101203; 010-207-1613/G07001020716130/ 103007, 105403, 124102, 127706; 010-213-1206/G07001021312060/ 105002, 105404, 121202, 124103, 127707, 101204; 010-213-1408/G07001021314080/ 102704, 105405, 121203, 124104, 127708; 010-213-1610/G07001021316100/ 103008, 105406, 130101; 010-213-1812/G07001021318120/ 103009, 105407, 127709; 010-219-1406/G07001021914060/ 102705, 105408, 117503, 119301, 121201, 124105; 010-219-1608/G07001021916080/ 103010, 105409, 121204, 127710; 010-219-1810/G07001021918100/ 103011, 105410, 127711; 010-225-1807/G07001022518070/ 102706, 113701, 117001, 121205; 010-225-2009/G07001022520090/ 103001, 117903, 127004, 127712; 010-307-1006/G07001030710060/ 102707, 106202, 118004, 129302; 010-307-1006-2/G070010307100620/ 117512, 127718, 114304; 010-307-1208/G07001030712080/ 102709, 105601, 117504; 010-307-1410/G07001030714100/ 102710, 105602, 127713; 010-307-1612/G07001030716120/ 103012, 105603, 117505; 010-313-1205/G07001031312050/ 103013, 105604, 117506, 129303; 010-313-1407/G07001031314070/ 102711, 105605, 117507, 124107, 128702, 130102; 010-313-1609/G07001031316090/ 103014, 105606; 010-313-1811/G07001031318110/ 103015, 105607, 127714; 010-313-2013/G07001031320130/ 103016, 105608, 127715; 010-319-1606/G07001031916060/ 102712, 105609, 114905, 117508; 010-319-1808/G07001031918080/ 103017, 105610, 125103; 010-319-2010/G07001031920100/ 103018, 105611, 125104; 010-325-2007/G07001032520070/ 102713, 105612, 121206; 010-325-2209/G07001032522090/ 105613, 121207, 101206, 101207, 101208. A RECORDED FDA enforcement action (occurred=true).",
      "claimType": "official_fact",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": true,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "SourceCitation",
        "source": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-2213-2026&lang=eng",
        "value": "Devices recall, status Ongoing",
        "adjudicator": {
          "name": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-2213-2026&lang=eng"
        }
      },
      "falsifier": "A RECORD of measured/reported events (occurred=true) from openFDA drug & device enforcement (recall) reports — U.S. FDA; superseded when openFDA drug & device enforcement (recall) reports — U.S. FDA publishes its next release."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-2213-2026/fda-classification",
      "@type": "schema:Claim",
      "name": "FDA classified recall Z-2213-2026 as Class II",
      "text": "The FDA assigned this recall Class II: Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.. This is the FDA's official risk classification — its regulatory DETERMINATION of the hazard category (a verdict constituted by the FDA's recognition, occurred=FALSE), adjudged=TRUE because the FDA issued it. It is NOT a measurement of harm to any specific person.",
      "claimType": "regulatory_determination",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "AdjudicativeAct",
        "source": "U.S. FDA recall classification (openFDA enforcement)",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-2213-2026&lang=eng",
        "value": "Class II",
        "adjudicator": {
          "name": "U.S. FDA recall classification (openFDA enforcement)",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-2213-2026&lang=eng"
        }
      },
      "falsifier": "occurred=FALSE: a regulatory DETERMINATION (an authority's classification/judgment) has NO material truthmaker independent of recognition — it is constituted by U.S. FDA recall classification (openFDA enforcement)'s say-so (a Searle institutional fact), not measured. adjudged=TRUE because U.S. FDA recall classification (openFDA enforcement) (an external authority) issued it. Distinct from a recorded measurement (which has an independent truthmaker) and from a modeled figure (a computed quantity, not a verdict). Challengeable; superseded if the authority reclassifies."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-2213-2026/boundary",
      "@type": "schema:Claim",
      "name": "What an FDA recall is — and is NOT",
      "text": "A recall is a RECORDED FDA enforcement action — the FDA logged that a firm is removing or correcting a product, classed by the FDA's risk judgment. It does NOT establish that any specific unit is affected or caused harm; causation is not established. This record informs and compares — it is NOT medical advice and does not tell anyone to start or stop a drug or device. For a medical decision, route to the FDA recall record and a licensed pharmacist or prescriber.",
      "claimType": "interpretation",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": false,
        "operative": true
      },
      "grounding": {
        "groundingType": "Interpretation",
        "source": "Atlas Commons L5 boundary (enforcement-action ≠ harm causation)",
        "sourceUrl": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
        "value": "a recorded enforcement action, not proof of harm or causation; not medical advice",
        "adjudicator": null
      },
      "falsifier": "occurred=FALSE (a framing, not an event) and adjudged=FALSE (an in-house reading, not an authority's verdict) — the LOWEST-warrant claim here. A flat representation would render this identically to the official facts; the four-status does not."
    }
  ],
  "hasPart": [
    {
      "@type": "schema:ItemList",
      "name": "FDA recall details — Z-2213-2026",
      "numberOfItems": 9,
      "itemListElement": [
        {
          "@type": "schema:ListItem",
          "position": 1,
          "item": {
            "Field": "Recall number",
            "Value": "Z-2213-2026",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 2,
          "item": {
            "Field": "Product type",
            "Value": "Devices",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 3,
          "item": {
            "Field": "Classification",
            "Value": "Class II",
            "Status": "determination"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 4,
          "item": {
            "Field": "Status",
            "Value": "Ongoing — the recall is still being carried out.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 5,
          "item": {
            "Field": "Recalling firm",
            "Value": "Novapproach Spine, LLC",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 6,
          "item": {
            "Field": "Reason",
            "Value": "Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 7,
          "item": {
            "Field": "Initiated",
            "Value": "2026-04-23",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 8,
          "item": {
            "Field": "Lot/code info",
            "Value": "REF/UDI-DI/ Cage Lots: 010-107-1007/G07001010710070/ 103003, 105701, 118001, 129301; 010-107-1007-2/G070010107100720/ 114301, 117509, 128703; 010-107-1209/G07001010712090/ 102714, 105702, 117501; 010-107-1411/G07001010714110/ 103019, 105703; 010-107-1613/G07001010716130/ 103020, 105704, 127701; 010-113-1006/G07001011310060/ 103021, 105705, 118002, 129801; 010-113-1006-2/G070010113100620/ 117510, 127716, 114302; 010-113-1208/G07001011312080/ 103022, 105706, 117502, 126902; 010-113-1410/G07001011314100/ 102715, 105707, 127702; 010-113-1612/G07001011316120/ 103023, 105708, 127703; 010-119-1407/G07001011914070/ 103024, 105709, 125101, 128701; 010-119-1609/G07001011916090/ 102716, 105710, 127001; 010-119-1811/G07001011918110/ 103025, 105711, 127003; 010-125-1606/G07001012516060/ 102717, 117901, 127002; 010-125-1808/G07001012518080/ 103005, 117902, 127704; 010-207-1007/G07001020710070/ 102702, 106201, 118003, 125102; 010-207-1007-2/G070010207100720/ 117511, 124106, 127717, 114303; 010-207-1209/G07001020712090/ 102701, 105401, 124101, 127705, 101201; 010-207-1411/G07001020714110/ 105001, 105402, 101203; 010-207-1613/G07001020716130/ 103007, 105403, 124102, 127706; 010-213-1206/G07001021312060/ 105002, 105404, 121202, 124103, 127707, 101204; 010-213-1408/G07001021314080/ 102704, 105405, 121203, 124104, 127708; 010-213-1610/G07001021316100/ 103008, 105406, 130101; 010-213-1812/G07001021318120/ 103009, 105407, 127709; 010-219-1406/G07001021914060/ 102705, 105408, 117503, 119301, 121201, 124105; 010-219-1608/G07001021916080/ 103010, 105409, 121204, 127710; 010-219-1810/G07001021918100/ 103011, 105410, 127711; 010-225-1807/G07001022518070/ 102706, 113701, 117001, 121205; 010-225-2009/G07001022520090/ 103001, 117903, 127004, 127712; 010-307-1006/G07001030710060/ 102707, 106202, 118004, 129302; 010-307-1006-2/G070010307100620/ 117512, 127718, 114304; 010-307-1208/G07001030712080/ 102709, 105601, 117504; 010-307-1410/G07001030714100/ 102710, 105602, 127713; 010-307-1612/G07001030716120/ 103012, 105603, 117505; 010-313-1205/G07001031312050/ 103013, 105604, 117506, 129303; 010-313-1407/G07001031314070/ 102711, 105605, 117507, 124107, 128702, 130102; 010-313-1609/G07001031316090/ 103014, 105606; 010-313-1811/G07001031318110/ 103015, 105607, 127714; 010-313-2013/G07001031320130/ 103016, 105608, 127715; 010-319-1606/G07001031916060/ 102712, 105609, 114905, 117508; 010-319-1808/G07001031918080/ 103017, 105610, 125103; 010-319-2010/G07001031920100/ 103018, 105611, 125104; 010-325-2007/G07001032520070/ 102713, 105612, 121206; 010-325-2209/G07001032522090/ 105613, 121207, 101206, 101207, 101208",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 9,
          "item": {
            "Field": "Distribution pattern",
            "Value": "US Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.",
            "Status": "recorded"
          }
        }
      ]
    }
  ]
}