YAMATO PLUS IAB; intra-aortic balloon catheter and accessories used to provide counter pulsation therapy in the aorta. Product Description Part Number YAMATO PLUS-R 30cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0559-01 YAMATO PLUS-R 30cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0559-02 YAMATO PLUS-R 35cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0560-01 YAMATO PLUS-R 35cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0560-02 YAMATO PLUS-R 40cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0561-01 YAMATO PLUS-R 40cc IAB WITH ACCESSORIES AND RETRACTION KIT D684-00-0561-02 PACKAGED INSERTION KIT - YAMATO PLUS-R 7.5 Fr. 30/35/40 cc IABs D886-00-0019-25

Why it was recalled

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Product Description                                                                                       UDI YAMATO PLUS-R 30cc IAB WITH ACCESSORIES AND RETRACTION KIT 10607567107585 YAMATO PLUS-R 30cc IAB WITH ACCESSORIES AND RETRACTION KIT 10607567107592 YAMATO PLUS-R 35cc IAB WITH ACCESSORIES AND RETRACTION KIT 10607567107608 YAMATO PLUS-R 35cc IAB WITH ACCESSORIES AND RETRACTION KIT 10607567107615 YAMATO PLUS-R 40cc IAB WITH ACCESSORIES AND RETRACTION KIT 10607567107622 YAMATO PLUS-R 40cc IAB WITH ACCESSORIES AND RETRACTION KIT 10607567107639

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Domestic distribution nationwide. International distribution worldwide.

Recall record

Recalling firm
Datascope Corp.
Product type
Devices
Recall number
Z-2216-2023
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
N/A
Origin
Fairfield, NJ
Recall initiated
2023-06-23

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2216-2023 · Source: openFDA enforcement.