Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) ANESTHESIA TRAY, Medline SKU DYNJRA1355A; 2) SPINAL ANESTHESIA TRAY, Medline SKU DYNJRA2151.
Why it was recalled
Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
Medline SKU DYNJRA1355A: UDI/DI 10195327179830 each, FI case, Lot Numbers: 26CBA272; Medline SKU DYNJRA1355A: UDI/DI 10195327179830 each, FI case, Lot Numbers: 26ABS307; Medline SKU DYNJRA1355A: UDI/DI 10195327179830 each, FI case, Lot Numbers: 25HBE663; Medline SKU DYNJRA1355A: UDI/DI 10195327179830 each, FI case, Lot Numbers: 25FBA408; Medline SKU DYNJRA1355A: UDI/DI 10195327179830 each, FI case, Lot Numbers: 25CBG043; Medline SKU DYNJRA1355A: UDI/DI 10195327179830 each, FI case, Lot Numbers: 24FBT395; Medline SKU DYNJRA2151: UDI/DI 10195327365400 each, FI case, Lot Numbers: 24FBT396.
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide - US Nationwide distribution in the US Territory of Virgin Islands and the countries of Bahamas, Panama, Barbados.
Recall record
- Recalling firm
- Medline Industries, LP
- Product type
- Devices
- Recall number
- Z-2231-2026
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Origin
- Northfield, IL
- Recall initiated
- 2026-04-10
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-2231-2026 · Source: openFDA enforcement.