Drug & Device Safety Atlas
Class II FDA device recall · Z-2256-2023

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical Sensor Assembly: (a) Symbia Pro.specta Q3, Model 11364751; (b) Symbia Pro.specta X3, Model 11364752; (c) Symbia Pro.specta X7, Model 11364753.

Why it was recalled

During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the patient than needed leading to a reduction in image resolution.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Software version VA20A.  (a) Serial numbers 100201, 100202, 100203, 100204, 100205, 100206, 100207, 100209, 100208, 100210, 100212, 100211, 100213, 100215, 100217, 100219, 100218, 100220, 100221, and 100224; UDI-DI 4056869264745.  Additional serial numbers as of 09/07/2023: 100223, 100225, 100227, 100228, 100237; Additional serial numbers as of 10/17/2023: 100214, 100229; Additional serial numbers as of 12/12/2023:  100232, 100231, 100234; Additional serial number as of 01/15/2024:100226.   (b) Serial numbers 100101, 100102, 100201, 100203, 100204, 100207, 100206, 100208, 100205, 100209, 100210, 100212, 100213, 100214, 100215, 100216, 100218, 100217, 100219, 100220, 100221, 100223, 100224, 100222, 100226, 100228, 100227, 100225, 100229, 100231, 100232, 100233, 100235, 100234, 100237, 100239, 100241, 100240, 100236, 100238, 100242, 100243, 100247, 100244, 100250, 100251, 100252, and 100255; UDI-DI 4056869264752;  Additional serial numbers as of 09/07/2023: 100245, 100256, 100257, 100258, 100261, 100262, 100265, 100273; Additional serial numbers as of 10/17/2023: 100230, 100266; Additional serial numbers as of 12/12/2023:  100269; Additional serial numbers as of 01/15/2024: 100248, 100254, 100260, 100271, 100272.   (c) Serial numbers 100101, 100201, 100203, 100202, 100205, and 100206; UDI-DI 4056869264769; Additional serial numbers as of 09/07/2023: 100208, 100210; Additional serial numbers as of 12/12/2023:  100209

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Distribution was made to ID, KS, MA, NH, OH, TX, and VA. There was no government/military distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Canada, Colombia, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, Netherlands, Portugal, Singapore, Switzerland, Turkey, United Arab Emirates, and United Kingdom.

Recall record

Recalling firm
Siemens Medical Solutions USA, Inc.
Product type
Devices
Recall number
Z-2256-2023
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Hoffman Estates, IL
Recall initiated
2023-07-12

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2256-2023 · Source: openFDA enforcement.