Class II FDA device recall · Z-2256-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

BD Pyxis, Name/REF: BD PYXIS MB TOWER MAIN/ 169-153, BD PYXIS MB TOWER MAIN CR/ 169-154, BD PYXIS MB TOWER MAIN W/MP/ 169-157, BD PYXIS MB TWR MN 1FM-10HH-1FH MP/ 169-143, BD PYXIS MB TWR MN 2FH-1FM-4HH-2FM MP/ 169-145, BD PYXIS MB TWR MN 2FM-8HH-1FH MP/ 169-144, BD PYXIS MB TWR MN 4FH-4HH-1FM MP/ 169-142, BD PYXIS MB TWR RXNOW MAIN W/MED PRINTER / 169-155, BD PYXIS MB TWR RXNOW MAIN W/PAT PRINTER/ 169-156, BD PYXIS MDBANK TWR MN 7HH-1HM-3FM/ 169-132, BD PYXIS MDBNK TWR MN CR-4HH-5FM-P RxNow/ 169-131, BD PYXIS MEDBANK TWR MN 10HH-1FH-1FM/ 169-23, BD PYXIS MEDBANK TWR MN 10HH-2FH/ 169-24, BD PYXIS MEDBANK TWR MN 10HH-2FM/ 169-25, BD PYXIS MEDBANK TWR MN 12HH-1FF/ 169-26, BD PYXIS MEDBANK TWR MN 12HH-1FH/ 169-27, BD PYXIS MEDBANK TWR MN 12HH-1FM/ 169-28, BD PYXIS MEDBANK TWR MN 14HH/ 169-29, BD PYXIS MEDBANK TWR MN 1FH-10HM-1FM/ 169-138, BD PYXIS MEDBANK TWR MN 1HH-1HM-2FM-4FF/ 169-30, BD PYXIS MEDBANK TWR MN 2FH-2HM-4FM/ 169-31, BD PYXIS MEDBANK TWR MN 2HH-10HM-1FM/ 169-32, BD PYXIS MEDBANK TWR MN 2HH-1FH-5FM/ 169-33, BD PYXIS MEDBANK TWR MN 2HH-2FH-4FM/ 169-34, BD PYXIS MEDBANK TWR MN 2HH-2FM-4FF/ 169-35, BD PYXIS MEDBANK TWR MN 2HH-4FH-2FM/ 169-139, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FF/ 169-36, BD PYXIS MEDBANK TWR MN 2HH-4HM-4FM/ 169-37, BD PYXIS MEDBANK TWR MN 2HH-6FM/ 169-38, BD PYXIS MEDBANK TWR MN 2HH-6HM-3FM/ 169-39, BD PYXIS MEDBANK TWR MN 2HH-8HM-2FM/ 169-40, BD PYXIS MEDBANK TWR MN 2HM-6FM/ 169-41, BD PYXIS MEDBANK TWR MN 3FH-4FM/ 169-42, BD PYXIS MEDBANK TWR MN 3FH-8HH MEDPASS/ 169-140, BD PYXIS MEDBANK TWR MN 3HH-7HM-2FM/ 169-43, BD PYXIS MEDBANK TWR MN 4HH-10HM/ 169-44, BD PYXIS MEDBANK TWR MN 4HH-1FH-2HM-3FM/ 169-45, BD PYXIS MEDBANK TWR MN 4HH-1FH-4FM/ 169-46, BD PYXIS MEDBANK TWR MN 4HH-1FH-4HM-2FM/ 169-47, BD PYXIS MEDBANK TWR MN 4HH-1FH-6HM-1FM/ 169-48, BD PYXIS MEDBANK TWR MN 4HH-2FH-2HM-2FM/ 169-49, BD PYXIS MEDBANK TWR MN 4HH-2FH-3FM/ 169-50, BD PYXIS MEDBANK TWR MN 4HH-2HM-4FM/ 169-51, BD PYXIS MEDBANK TWR MN 4HH-3FH-2FM/ 169-52, BD PYXIS MEDBANK TWR MN 4HH-4FH-1FM/ 169-53, BD PYXIS MEDBANK TWR MN 4HH-4HM-3FM/ 169-54, BD PYXIS MEDBANK TWR MN 4HH-5FH / 169-55, BD PYXIS MEDBANK TWR MN 4HH-5FM/ 169-56, BD PYXIS MEDBANK TWR MN 4HH-6HM-2FM/ 169-57, BD PYXIS MEDBANK TWR MN 4HH-8HM-1FM/ 169-58, BD PYXIS MEDBANK TWR MN 4HM-5FM/ 169-59, BD PYXIS MEDBANK TWR MN 5HH-3HM-3FM/ 169-60, BD PYXIS MEDBANK TWR MN 5HH-9HM/ 169-61, BD PYXIS MEDBANK TWR MN 6FH-1FM/ 169-62, BD PYXIS MEDBANK TWR MN 6HH-1FH-2HM-2FM/ 169-63, BD PYXIS MEDBANK TWR MN 6HH-1FH-3FM/ 169-64, BD PYXIS MEDBANK TWR MN 6HH-2FH-2FM/ 169-65, BD PYXIS MEDBANK TWR MN 6HH-2FH-2HM-1FM/ 169-66, BD PYXIS MEDBANK TWR MN 6HH-4FH / 169-67, BD PYXIS MEDBANK TWR MN 6HH-4FM/ 169-68, BD PYXIS MEDBANK TWR MN 6HH-4HM-2FM/ 169-69, BD PYXIS MEDBANK TWR MN 6HH-6HM-1FM/ 169-70, BD PYXIS MEDBANK TWR MN 6HH-8HM/ 169-71, BD PYXIS MEDBANK TWR MN 7FH/ 169-72, BD PYXIS MEDBANK TWR MN 7FM/ 169-73, BD PYXIS MEDBANK TWR MN 7HH-1HM-3FM/ 169-74, BD PYXIS MEDBANK TWR MN 8HH-1FH-2FM/ 169-75, BD PYXIS MEDBANK TWR MN 8HH-1FH-2HM-1FM/ 169-76, BD PYXIS MEDBANK TWR MN 8HH-2FH-1FM/ 169-77, BD PYXIS MEDBANK TWR MN 8HH-2HM-2FM/ 169-78, BD PYXIS MEDBANK TWR MN 8HH-3FF/ 169-79, BD PYXIS MEDBANK TWR MN 8HH-3FH/ 169-80, BD PYXIS MEDBANK TWR MN 8HH-3FM/ 169-81, BD PYXIS MEDBANK TWR MN CR-10HH-2FH-P/ 169-82, BD PYXIS MEDBANK TWR MN CR-10HH-2FM-P/ 169-83, BD PYXIS MEDBANK TWR MN CR-12HH-1FH/ 169-84, BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P/ 169-85, BD PYXIS MEDBANK TWR MN CR-14HH-P/ 169-86, BD PYXIS MEDBANK TWR MN CR-2HH-1FH-5FM/ 169-87, BD PYXIS MEDBANK TWR MN CR-2HH-6FH/ 169-88, BD PYXIS MEDBANK TWR MN CR-2HH-6FM/ 169-89, BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P/ 169-90, BD PYXIS MEDBANK TWR MN CR-4HH-5FM-P/ 169-91, BD PYXIS MEDBANK TWR MN CR-6HH-1FH-3FM-P/ 169-92, BD PYXIS MEDBANK TWR MN CR-6HH-2FH-2FM-P/ 169-93, BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P/ 169-94, BD PYXIS MEDBANK TWR MN CR-6HH-8HM-P/ 169-95, BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P/ 169-96, BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P/ 169-97, BD

Why it was recalled

Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI-DI: 10885403512568. Devices Manufactured November 15, 2012 - July, 25 2023  User Guide, DIR 10000414989/DIR 10000414987/DIR 10000414990, Version 00.

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide distribution including in the states of NY, CA, MO, AZ, CO, KS, IL, FL, OK, OH, TN, GA, NM, NV, NE, TX, NC, KY, WI, SC, IA, MT, SD, VA, NJ, PA, AK, MS, MI, WA, UT, ND, MA, MN, OR, IN, WV, AL, ID, LA, CT, WY, MD, NH, AR, HI, DC.

Recall record

Recalling firm
CareFusion 303, Inc.
Product type
Devices
Recall number
Z-2256-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
San Diego, CA
Recall initiated
2024-05-13

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2256-2024 · Source: openFDA enforcement.