ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). Model/Catalog Number: G5S-00A G5S-00A-TSO G5S-00C G5S-01A G5S-02A G5S-02C G5S-02-L G5S-04A G5S-05A G5S-05C G5S-06A G5S-06C G5S-08A G5S-08C G5S-10A G5S-10C G5S-11A G5S-11C G5S-12C G5S-13A G5S-14A G5S-15A G5S-17A G5S-17C G5S-19A G5S-23C G5S-29A G5S-29C G5S-31A G5S-31A-SJA G5S-31C-SJA G5S-36A G5S-37A G5S-41A G5S-41C G5S-80A G5S-80A-TSO G5S-80C G5S-80-L G5S-82A G5S-82C G5S-83C G5S-90C

Why it was recalled

G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
GTIN: 00812394021239 00812394021277 00812394021284 00812394021413 00812394022779 00847946038182 00847946038199 00847946038243 00847946038250 00847946038304 00847946038328 00847946038335 00847946038342 00847946038359 00847946038380 00847946038397 00847946038403 00847946038410 00847946038434 00847946038441 00847946038496 00847946038519 00847946038533 00847946038571 00847946038588 00847946038663 00847946038816 00847946038830 00847946039028 00847946039035 00847946039066 00847946039127     Serial Number Range: D00000276194-D00000343449

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Nationwide including PR. Foreign: AU BM CA CH DE DK ES FR GB HK HU ID IE IL IT KR MA MT NL NO PR SA SE SI SN TR

Recall record

Recalling firm
Zoll Medical Corporation
Product type
Devices
Recall number
Z-2324-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
E-Mail
Origin
Deerfield, WI
Recall initiated
2024-05-31

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.