Class I FDA device recall · Z-2446-2024

Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.

Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB560 VENTILATOR EU-DIV/4096600-01, PURITAN BENNETT 560 VENTILATOR JAPAN/4096600-02, PURITAN BENNETT 560 VENTILATOR APAC/4096600-03, Puritan Bennett 560 Ventilator LATAM/4096600-04, Puritan Bennett 560 Ventilator EMEA/4096600-05, Puritan Bennett 560 Ventilator CAN ANZ/4096600-06, Puritan Bennett 540 Ventilator USA/4097100 Puritan Bennett 520 Ventilator/4098300, Puritan Bennett 520 Ventilator EU-DIV/4098300-01, Puritan Bennett 520 Ventilator Japan/4098300-02, Puritan Bennett 520 Ventilator APAC/4098300-03, Puritan Bennett 520 Ventilator EMEA/4098300-05, Puritan Bennett 520 Ventilator CAN/ANZ/4098300-06, DL4096600 DL PB560 VENTILATOR/DL4096600, DL4096600-06 RFB PB560 VENT CAN ANZ/DL4096600-06, REFURBISHED PB540 VENTILATOR/DL4097100, Puritan Bennett 540 Ventilator Refurbished/DS4097100

Why it was recalled

Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
REF/UDI-DI(GTIN): 4096600/10884521087798 4096600-01/10884521195998 4096600-02/10884521196001 4096600-03/10884521196018 4096600-04/10884521196025 4096600-05/10884521196032 4096600-06/10884521196377 4097100/4097100 4098300/10884521183636 4098300-01/10884521195943 4098300-02/4098300-02 4098300-03/10884521195967 4098300-05/10884521195981 4098300-06/10884521196384 DL4096600/10884521786509 DL4096600-06/10884521786516 DL4097100/DL4097100 DS4097100/DS4097100  All device serial numbers used before distribution of PB500 Series Ventilator User Manual and Clinician's Manual addendum, PT00196729 A, 2024-06-11

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US: KY, OH, GA, MN, IN,UT, IL, NJ, NY, PR, LA, MD, MO, TN, MA, ME, CA, SC, AR, TX, WI, PA, CO, MI, NM, RI, VA, DC, VT, NC, NH, WV, NV, IA. OUS: Argentina, Australia, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Lithuania, Malaysia, Mexico, Montenegro, Namibia, Nepal, New Zealand, Nigeria, Panama, Poland, Reunion, Romania, Russian Federation, Saudi Arabia, Slovakia, Slovenia, Spain, Sweden, Taiwan, Turkey, Ukraine, United Kingdom, Viet Nam, South Africa, Austria, Kenya, Latvia, Lebanon, Mongolia, Nicaragua, Thailand, Tunisia, Algeria, Jordan, Libya, Malawi, Qatar, Pakistan, United Arab Emirates, Oman, Peru, Philippines, Bhutan, Iran - Islamic Republic Of, Bolivia, Kuwait, Egypt, Switzerland, Sudan, Norway, Korea - Republic Of, Bahrain, Ethiopia, Paraguay

Recall record

Recalling firm
Covidien
Product type
Devices
Recall number
Z-2446-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Carlsbad, CA
Recall initiated
2024-06-24

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2446-2024 · Source: openFDA enforcement.