Drug & Device Safety Atlas
Class II FDA device recall · Z-2448-2023

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgM, Prealbumin, RF and Transferrin assays Catalogue Number: IT2691

Why it was recalled

As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
GTIN: 05055273204032 Batch/  Lot number Expiry Date   590858  28 08 23;   590859  28 08 23;   627222  28Jul 24;   627224  28Jul24;   634886  28Jul24;   634887  28Jul24

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

CA, FL, IL, ME, MI, NC, OH, PA,NJ,

Recall record

Recalling firm
Randox Laboratories Ltd.
Product type
Devices
Recall number
Z-2448-2023
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
E-Mail
Origin
Crumlin (North), N/A
Recall initiated
2023-07-19

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2448-2023 · Source: openFDA enforcement.