Atellica CH Gamma-Glutamyl Transferase-IIn vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma (lithium heparin Test Code: GGT Siemens Material Number (SMN): 11097597

Why it was recalled

Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿CH Total Bilirubin_2 (TBil_2), Atellica¿ CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica¿ CH HDL Cholesterol (HDLC)

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Unique Device Identification (UDI): 00630414596440 All Lot Numbers

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam

Recall record

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product type
Devices
Recall number
Z-2473-2023
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Tarrytown, NY
Recall initiated
2023-07-10

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2473-2023 · Source: openFDA enforcement.