Drug & Device Safety Atlas
Class II FDA device recall · Z-2481-2023

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee prosthesis implant during a total knee arthroplasty surgical procedure Item Number: 17-5220/01

Why it was recalled

A crack in the weld of the rasp handle may lead to a fracture of the frontal sleeve when the rasp is further impacted or extracted. The rasp remains intact. The rasp would need to be pulled out by a clamp leading to a prolongation of surgery or modified surgery

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI-DI: 04026575284658 Lot Codes: C010108, C151519

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

IN,GA, TX

Recall record

Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Product type
Devices
Recall number
Z-2481-2023
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
E-Mail
Origin
Norderstedt, N/A
Recall initiated
2023-07-26

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2481-2023 · Source: openFDA enforcement.