Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.
Proclaim Plus 7 Implantable Pulse Generator REF 3672 Product Description: The implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG can deliver stimulation with a single program or with multiple programs. Each program can provide stimulation to a single anatomical area or to multiple areas. The IPG communicates wirelessly with system programmers and controllers, and the IPGs are available in small and large sizes to accommodate different power needs. The Proclaim" XR and Proclaim" Plus IPGs are used for Spinal Cord Stimulation (SCS). The Proclaim" DRG IPG is used for Dorsal Root Ganglion (DRG) stimulation. The Infinity" IPG is used for Deep Brain Stimulation (DBS).
Why it was recalled
Firm has received complaints from patients who are unable to exit MRI (Magnetic Resonance Imaging) mode on their IPGs (Implantable Pulse Generators).
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
Model Number: 3672 UDI-DI Codes: 05415067046406 Serial Numbers: US CBW328.1 CCA167.1 CCA177.1 CCU613.1 CCV760.1 CCX700.1 CCY890.1 CDR307.1 CDW220.1 CEC548.1 CEC719.1 CEE095.1 CEH111.1 CEH156.1 CEH200.1 CEH272.1 CEH293.1 CEH334.1 CEH964.1 CEH975.1 CEH993.1 CEK127.1 CEK258.1 CEK265.1 CEK271.1 CEK310.1 CEK432.1 CEM232.1 CEM340.1 CEM349.1 CEM375.1 CEM383.1 CEM396.1 CEN429.1 CEN444.1 CEP494.1 CEP497.1 CEP537.1 CEP540.1 CEP541.1 CEP553.1 CEP554.1 CEP562.1 CEP585.1 CEP645.1 CEP658.1 CEP663.1 CEP671.1 CEP685.1 CEP725.1 CEP730.1 CEP838.1 CEP847.1 CEP867.1 CER393.1 CER395.1 CER462.1 CER482.1 CER509.1 CER576.1 CET134.1 CET147.1 CET298.1 CET336.1 CET343.1 CET649.1 CEY608.1 CEY648.1 CFL299.1 CFL403.1 CFL540.1 CFL582.1 CFP963.1 CFP987.1 CFP991.1 CFR079.1 CFR100.1 CFR121.1 CFR392.1 CFR446.1 CFR478.1 CFR585.1 CFU079.1 CFU082.1 CFU091.1 CFU097.1 CFU098.1 CFU099.1 CFU121.1 CFU126.1 CFU132.1 CFU145.1 CFU333.1 CFU401.1 CFU426.1 CFU441.1 CFU460.1 CFU467.1 CFU472.1 CFU483.1 CFU526.1 CFU547.1 CFU569.1 CFU646.1 CFU687.1 CFV457.1 CFV588.1 CFV718.1 CFX211.1 CGA013.1 CGA029.1 CGA065.1 CGA097.1 CGA145.1 CGA331.1 CGA365.1 CGA406.1 CGA426.1 CGA451.1 CGA491.1 CGA500.1 CGA515.1 CGA581.1 CGC898.1 CGC908.1 CGC916.1 CGC921.1 CGC926.1 CGC955.1 CGC957.1 CGD063.1 CGD070.1 CGD076.1 CGD085.1 CGD101.1 CGD111.1 CGD119.1 CGD123.1 CGD192.1 CGD193.1 CGD220.1 CGD231.1 CGD240.1 CGD248.1 CGD251.1 CGD277.1 CGD278.1 CGD283.1 CGD308.1 CGD342.1 CGD343.1 CGD362.1 CGD694.1 CGD719.1 CGD739.1 CGD777.1 CGD790.1 CGD832.1 CGD892.1 CGG890.1 CGH131.1 CGH175.1 CGH183.1 CGH199.1 CGH245.1 CGH346.1 CGH413.1 CGL108.1 CGL207.1 CGL321.1 CGL863.1 CGL866.1 CGL889.1 CGP051.1 CGP078.1 CGP326.1 CGP615.1 CGS914.1 CGW231.1 CHB011.1 CHE598.1 CHE845.1 CHF161.1 CHF178.1 CHH717.1 CHH744.1 CHH773.1 CHJ032.1 CHK331.1 CHK354.1 CHK434.1 CHK648.1 CHM211.1 CHM389.1 CHM509.1 CHP603.1 CHP605.1 CHP621.1 CHR155.1 CHR240.1 CHT217.1 CHT238.1 CHT240.1 CHT241.1 CHT243.1 CHT245.1 CHT272.1 CHT356.1 CHT361.1 CHT373.1 CHT469.1 CHT478.1 CHT539.1 CHT553.1 CHT561.1 CHT589.1 CHT609.1 CHT624.1 CHT758.1 CHT815.1 CHT834.1 CHV674.1 CHV680.1 CHV710.1 CHV713.1 CHV794.1 CHV901.1 CHV989.1 CHW217.1 CHW262.1 CHW263.1 CHW306.1 CHW350.1 CHW362.1 CHW526.1 CHW560.1 CHW590.1 CHW591.1 CHW640.1 CHZ471.1 CHZ563.1 CHZ576.1 CJA246.1 CJG129.1 CJG177.1 CJM932.1 CJM940.1 CJN773.1 CJN780.1 CJN879.1 CJN987.1 CJP047.1 CJW418.1 CJW475.1 CJW534.1 CJW548.1 CJW586.1 CJX925.1 CJX962.1 CJX982.1 CJX986.1 CJY012.1 CJY075.1 CJY111.1 CJY226.1 CJY238.1 CJY304.1 CJY323.1 CJY360.1 CJY409.1 CJZ061.1 CJZ103.1 CJZ135.1 CJZ157.1 CJZ224.1 CJZ241.1 CJZ253.1 CJZ256.1 CJZ283.1 CJZ293.1 CJZ300.1 CJZ330.1 CJZ420.1 CJZ444.1 CJZ450.1 CKB402.1 CKB428.1 CKB439.1 CKB447.1 CKB487.1 CKB507.1 CKB608.1 CKB624.1 CKB654.1 CKB696.1 CKB725.1 CKB767.1 CKB887.1 CKD072.1 CKF569.1 CKF587.1 CKG394.1 CKG550.1 CKG586.1 CKG607.1 CKG915.1 CKG954.1 CKG956.1 CKG962.1 CKH467.1 CKH606.1 CKH692.1 CKH719.1 CKH736.1 CKH765.1 CKH774.1 CKH786.1 CKH815.1 CKH859.1 CKH923.1 CKL195.1 CKL217.1 CKL236.1 CKL295.1 CKL345.1 CKL464.1 CKL487.1 CKP251.1 CKP255.1 CKP277.1 CKP291.1 CKP393.1 CKP435.1 CKP483.1 CKP495.1 CKP563.1 CKP656.1 CKP678.1 CKP682.1 CKP717.1 CKP733.1 CKP803.1 CKP939.1 CKS483.1 CKS519.1 CKS531.1 CKS552.1 CKS588.1 CKS590.1 CKS616.1 CKS641.1 CKS662.1 CKS673.1 CKS684.1 CKS692.1 CKS698.1 CKS706.1 CKS827.1 CKS856.1 CKS869.1 CKS954.1 CKS986.1 CKS992.1 CKS994.1 CKS997.1 CKT159.1 CKY723.1 CKY781.1 CLC604.1 CLD150.1 CLD293.1 CLF739.1 CLF855.1 CLF880.1 CLF888.1 CLF901.1 CLG005.1 CLG094.1 CLG124.1 CLG391.1 CLG393.1 CLG395.1 CLG396.1 CLG423.1 CLG457.1 CLG534.1 CLK164.1
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Korean, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.
Recall record
- Recalling firm
- Abbott Medical
- Product type
- Devices
- Recall number
- Z-2508-2023
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Plano, TX
- Recall initiated
- 2023-07-18
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.