Class II FDA device recall · Z-2583-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately: HeartMate 3 LVAS Implant Kit, US, Model: 106524US; HeartMate 3 LVAS Implant Kit, OUS, Model: 106524INT; HeartMate 3 System Controller, US, Model: 106531US; HeartMate 3 System Controller, OUS, Model: 106531INT; HeartMate 3 System Controller Low Flow 2.0, Global, Model: 106531LF2

Why it was recalled

Left ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies, which could result in unexpected and false positive alarms, loss of visual alarms, loss of user interface(use of buttons, loss of visual alarms,etc.), loss of power, and pump stop

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Model/UDI-DI/System Controller Serial Numbers:   106524US/00813024013297/HSC-134168, HSC-134392, HSC-134393, HSC-134394, HSC-134395, HSC-134396, HSC-134397, HSC-134398, HSC-134399, HSC-134400, HSC-134401, HSC-134402, HSC-134403, HSC-134404, HSC-134405, HSC-134406, HSC-134407, HSC-134408, HSC-134409, HSC-134410, HSC-134411, HSC-134432, HSC-134433, HSC-134434, HSC-134436, HSC-134437, HSC-134438, HSC-134440, HSC-134452, HSC-134454, HSC-134502, HSC-134503, HSC-134504, HSC-134505, HSC-134506, HSC-134507, HSC-134508, HSC-134509, HSC-134510, HSC-134511, HSC-134512, HSC-134513, HSC-134514, HSC-134515, HSC-134516, HSC-134517, HSC-134518, HSC-134519, HSC-134520, HSC-134521, HSC-134572, HSC-134573, HSC-134574, HSC-134575, HSC-134576, HSC-134577, HSC-134578, HSC-134579, HSC-134580, HSC-134581, HSC-134582, HSC-134583, HSC-134584, HSC-134585, HSC-134586, HSC-134587, HSC-134588, HSC-134589, HSC-134590, HSC-134591, HSC-134602, HSC-134603, HSC-134604, HSC-134605, HSC-134606, HSC-134607, HSC-134608, HSC-134609, HSC-134610, HSC-134611, HSC-134612, HSC-134613, HSC-134614, HSC-134615, HSC-134616, HSC-134617, HSC-134618, HSC-134619, HSC-134620, HSC-134621, HSC-134662, HSC-134663, HSC-134664, HSC-134665, HSC-134666, HSC-134667, HSC-134668, HSC-134669, HSC-134670, HSC-134671, HSC-134672, HSC-134673, HSC-134674, HSC-134675, HSC-134676, HSC-134677, HSC-134678, HSC-134679, HSC-134680, HSC-134748, HSC-134749, HSC-134750, HSC-134806, HSC-134807, HSC-134808, HSC-134809, HSC-134810, HSC-134811, HSC-134812, HSC-134813, HSC-134814, HSC-134815, HSC-134816, HSC-134818, HSC-134819, HSC-134820, HSC-134821, HSC-134822, HSC-134823, HSC-134824, HSC-134825, HSC-134826, HSC-134827, HSC-134828, HSC-134829, HSC-134830, HSC-134831, HSC-134832, HSC-134833, HSC-134834, HSC-134835, HSC-134836, HSC-134837, HSC-134838, HSC-134839, HSC-134840, HSC-134841, HSC-134842, HSC-134843, HSC-134844, HSC-134845, HSC-134856, HSC-134857, HSC-134858, HSC-134859, HSC-134860, HSC-134861, HSC-134862, HSC-134863, HSC-134864, HSC-134865, HSC-134866, HSC-134867, HSC-134868, HSC-134869, HSC-134870, HSC-134871, HSC-134872, HSC-134873, HSC-134875, HSC-134876, HSC-134877, HSC-134878, HSC-134879, HSC-134880, HSC-134881, HSC-134882, HSC-134883, HSC-134885, HSC-134976, HSC-134977, HSC-134978, HSC-134979, HSC-134980, HSC-134981, HSC-134982, HSC-134983, HSC-134984.  106524INT/00813024011712/HSC-134262, HSC-134263, HSC-134264, HSC-134265, HSC-134266, HSC-134267, HSC-134268, HSC-134269, HSC-134270, HSC-134271, HSC-134272, HSC-134274, HSC-134275, HSC-134276, HSC-134277, HSC-134278, HSC-134279, HSC-134280, HSC-134281, HSC-134282, HSC-134283, HSC-134284, HSC-134285, HSC-134286, HSC-134287, HSC-134288, HSC-134289, HSC-134290, HSC-134291, HSC-134292, HSC-134293, HSC-134294, HSC-134295, HSC-134296, HSC-134297, HSC-134298, HSC-134299, HSC-134300, HSC-134301, HSC-134352, HSC-134353, HSC-134356, HSC-134357, HSC-134358, HSC-134359, HSC-134360, HSC-134361, HSC-134362, HSC-134364, HSC-134365, HSC-134368, HSC-134369, HSC-134371, HSC-134532, HSC-134533, HSC-134534, HSC-134535, HSC-134536, HSC-134537, HSC-134538, HSC-134539, HSC-134540, HSC-134541, HSC-134542, HSC-134543, HSC-134544, HSC-134545, HSC-134546, HSC-134547, HSC-134548, HSC-134549, HSC-134550, HSC-134551, HSC-134552, HSC-134553, HSC-134554, HSC-134555, HSC-134556, HSC-134557, HSC-134558, HSC-134559, HSC-134560, HSC-134561, HSC-134562, HSC-134563, HSC-134564, HSC-134565, HSC-134566, HSC-134567, HSC-134568, HSC-134569, HSC-134570, HSC-134571, HSC-134706, HSC-134707, HSC-134708, HSC-134709, HSC-134710, HSC-134711, HSC-134712, HSC-134713, HSC-134714, HSC-134715, HSC-134716, HSC-134717, HSC-134718, HSC-134719, HSC-134720, HSC-134721, HSC-134722, HSC-134723, HSC-134724, HSC-134725, HSC-134746, HSC-134747, HSC-134778, HSC-134779, HSC-134780, HSC-134781, HSC-134782, HSC-134783, HSC-134784, HSC-134785, HSC-134846, HSC-134848, HSC-134849, HSC-134850, HSC-134851, HSC-134852, HSC-134853, HSC-134854, HSC-134855, HSC-134957, HSC-134958, HSC-134959, HSC-134960, HSC-134961, HSC-134962, HSC-134964, HSC-134965, HSC-134966, HSC-134968, HSC-134969, HSC-134970, HSC-134971, HSC-134972, HSC-134973, HSC-134974, HSC-134975.  106531US/00813024013235/HSC-134242, HSC-134244, HSC-134245, HSC-134246, HSC-134247, HSC-134252, HSC-134253, HSC-134254, HSC-134255, HSC-134256, HSC-134257, HSC-134258, HSC-134259, HSC-134260, HSC-134261, HSC-134302, HSC-134303, HSC-134304, HSC-134305, HSC-134306, HSC-134307, HSC-134309, HSC-134310, HSC-134311, HSC-134312, HSC-134313, HSC-134314, HSC-134315, HSC-134316, HSC-134317, HSC-134318, HSC-134319, HSC-134320, HSC-134321, HSC-134322, HSC-134323, HSC-134324, HSC-134325, HSC-134326, HSC-134327, HSC-134328, HSC-134329, HSC-134330, HSC-134331, HSC-134332, HSC-134333, HSC-134334, HSC-134335, HSC-134336, HSC-134337, HSC-134338, HSC-134339, HSC-134340, HSC-134341, HSC-134342, HSC-134343, HSC-134344, HSC-134345, HSC-134346, HSC-134347, HSC-134348, HSC-134349, HSC-134350, HSC-134351, HSC-134372, HSC-134373, HSC-134374, HSC-134375, HSC-134376, HSC-134377, HSC-134378, HSC-134379, HSC-134380, HSC-134381, HSC-134382, HSC-134383, HSC-134384, HSC-134385, HSC-134386, HSC-134387, HSC-134388, HSC-134389, HSC-134390, HSC-134391, HSC-134412, HSC-134413, HSC-134414, HSC-134415, HSC-134416, HSC-134417, HSC-134418, HSC-134419, HSC-134420, HSC-134422, HSC-134423, HSC-134424, HSC-134425, HSC-134426, HSC-134427, HSC-134428, HSC-134429, HSC-134430, HSC-134431, HSC-134442, HSC-134443, HSC-134444, HSC-134445, HSC-134446, HSC-134447, HSC-134448, HSC-134449, HSC-134450, HSC-134451, HSC-134462, HSC-134463, HSC-134464, HSC-134465, HSC-134466, HSC-134467, HSC-134468, HSC-134469, HSC-134470, HSC-134471, HSC-134472, HSC-134473, HSC-134474, HSC-134475, HSC-134476, HSC-134477, HSC-134478, HSC-134479, HSC-134480, HSC-134481, HSC-134482, HSC-134483, HSC-134484, HSC-134485, HSC-134486, HSC-134487, HSC-134488, HSC-134489, HSC-134490, HSC-134491, HSC-134492, HSC-134493, HSC-134494, HSC-134495, HSC-134496, HSC-134497, HSC-134498, HSC-134499, HSC-134500, HSC-134501, HSC-134522, HSC-134523, HSC-134524, HSC-134525, HSC-134526, HSC-134527, HSC-134528, HSC-134529, HSC-134530, HSC-134531, HSC-134593, HSC-134622, HSC-134623, HSC-134624, HSC-134625, HSC-134626, HSC-134627, HSC-134628, HSC-134629, HSC-134630, HSC-134631, HSC-134632, HSC-134633, HSC-134634, HSC-134635, HSC-134636, HSC-134637, HSC-134638, HSC-134639, HSC-134640, HSC-134641, HSC-134642, HSC-134643, HSC-134644, HSC-134645, HSC-134646, HSC-134647, HSC-134648, HSC-134649, HSC-134650, HSC-134651, HSC-134652, HSC-134653, HSC-134654, HSC-134655, HSC-134656, HSC-134657, HSC-134658, HSC-134659, HSC-134660, HSC-134661, HSC-134682, HSC-134683, HSC-134684, HSC-134685, HSC-134686, HSC-134687, HSC-134688, HSC-134689, HSC-134690, HSC-134691, HSC-134692, HSC-134693, HSC-134694, HSC-134695, HSC-134696, HSC-134697, HSC-134698, HSC-134699, HSC-134700, HSC-134701, HSC-134730, HSC-134736, HSC-134737, HSC-134738, HSC-134739, HSC-134740, HSC-134741, HSC-134742, HSC-134743, HSC-134744, HSC-134745, HSC-134755, HSC-134756, HSC-134757, HSC-134758, HSC-134759, HSC-134760, HSC-134761, HSC-134762, HSC-134763, HSC-134764, HSC-134765, HSC-134766, HSC-134767, HSC-134768, HSC-134769, HSC-134770, HSC-134771, HSC-134772, HSC-134773, HSC-134774, HSC-134775, HSC-134776, HSC-134786, HSC-134787, HSC-134788, HSC-134789, HSC-134790, HSC-134791, HSC-134792, HSC-134793, HSC-134794, HSC-134795, HSC-134796, HSC-134797, HSC-134798, HSC-134799, HSC-134800, HSC-134801, HSC-134802, HSC-134803, HSC-134804, HSC-134805, HSC-134886, HSC-134887, HSC-134888, HSC-134889, HSC-134890, HSC-134891, HSC-134892, HSC-134893, HSC-134894, HSC-134895, HSC-134896, HSC-134897, HSC-134898, HSC-134899, HSC-134900, HSC-134901, HSC-134902, HSC-134903, HSC-134904, HSC-134905, HSC-134906, HSC-134907, HSC-134908, HSC-134909, HSC-134910, HSC-134911, HSC-134912, HSC-134913, HSC-134914, HSC-134915, HSC-134916, HSC-134917, HSC-134918, HSC-134919, HSC-134920, HSC-134921, HSC-134922, HSC-134923, HSC-134924, HSC-134925, HSC-134927, HSC-134933, HSC-134934, HSC-134936, HSC-134937, HSC-134938, HSC-134939, HSC-134940, HSC-134942, HSC-134943, HSC-134944, HSC-134945, HSC-134985, HSC-134996, HSC-134997, HSC-134998, HSC-134999.  106531INT/00813024011842/HSC-134946, HSC-134947, HSC-134948, HSC-134949, HSC-134951, HSC-134952, HSC-134954, HSC-134955, HSC-134986, HSC-134987, HSC-134988, HSC-134989, HSC-134990, HSC-134991, HSC-134992, HSC-134993, HSC-134994.  106531LF2/05415067031105/HSC-133326, HSC-134726, HSC-134729, HSC-134733, HSC-134735, HSC-134752

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US: MN, PA, FL, IL, NY, DE, AR, NJ, NC, GA, WI, AZ, TN, MO, TX, MA, OH, VA, NE, CA, OK, CO, DC, IN, MI, CT, UT, IA, MD, OR, NM, ME, SC, KY, LA, WA, NV, HI, AL, MS, KS, WV. OUS: Algeria, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia & Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan

Recall record

Recalling firm
Thoratec LLC
Product type
Devices
Recall number
Z-2583-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Origin
Pleasanton, CA
Recall initiated
2024-06-25

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2583-2024 · Source: openFDA enforcement.