Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Membrane Changer Single-Use, REF MC, each package contains 3 changers; Membrane Changer Reloadable, REF MC-R, each package contains 3 changers; and Membrane Changer Insert, REF MC-I, each package contains 5 inserts. All are components of the Sentec Digital Monitoring System (SDMS).
Why it was recalled
During rework of components during the manufacturing process, an additional piece of material may have been left in the device, which may result in an impact on measurement performance.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
REF MC: 230441, 230224, 230209, 230191, 230185, 230167, 230165, 230146, 230125, 230016, 225261, 225226, 225183, 225173, 225149, 225147, 225126, 225122, 225088, 225073, 225063, 225006; UDI-DI numbers 07640121880933, 17640121880930; REF MC-R: 230461, 230329, 230351, 230235, 230083, 230053, 230043, 230008, 230005, 225253, 225221; UDI-DI numbers 07640121880087, 17640121880084, 27640121880081; REF MC-I: 230372, 230339, 230336, 230229, 230206, 230132, 230117, 230051, 230044, 230012, 225299, 225298, 225280, 225255, 225218, 225216, 225190, 225185, 225184, 225175, 225120, 225091, 225089, 225064, 225058, 225033, 225023, 221064, 221056; UDI-DI numbers 07640121880506, 17640121880503. REVISED 9/21/2023: Additional lot number added for the U.S. for REF MC-I - 225119
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
US Nationwide distribution.
Recall record
- Recalling firm
- SenTec AG
- Product type
- Devices
- Recall number
- Z-2592-2023
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Therwil
- Recall initiated
- 2023-07-25
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-2592-2023 · Source: openFDA enforcement.