Class I FDA device recall · Z-2616-2025

Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.

Brand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable Product Description: The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates. The device can be used for mechanical ventilation, manual ventilation, pressure-supported spontaneous breathing, and spontaneous breathing. The device is specified for inhalational anesthesia and/or patient ventilation in accordance with the intended use during surgical or diagnostic interventions Component: Not applicable

Why it was recalled

The device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Lot Code: UDI-DI Number: 04048675556176 Serial Numbers Distributed in the US: ASTB-0181 ASTC-0005,ASTC-0006,ASTC-0007 ASTD-0244,ASTD-0259 ASTE-0180,ASTE-0254,ASTE-0255  ASTH-0083,ASTH-0464 ASUA-0479,ASUA-0482 ASUB-0012,ASUB-0013,ASUB-0015,ASUB-0017,ASUB-0018,ASUB-0292 ASUC-0065,ASUC-0084,ASUC-0085,ASUC-0086,ASUC-0087,ASUC-0088,ASUC-0090,ASUC-0091, ASUC-0092,ASUC-0096,ASUC-0097,ASUC-0099,ASUC-0100,ASUC-0101,ASUC-0102,ASUC-0103, ASUC-0104,ASUC-0105,ASUC-0107,ASUC-0108,ASUC-0109,ASUC-0110,ASUC-0111,ASUC-0112, ASUC-0113,ASUC-0116,ASUC-0117,ASUC-0118,ASUC-0119,ASUC-0120,ASUC-0121,ASUC-0122, ASUC-0123,ASUC-0125,ASUC-0126,ASUC-0127,ASUC-0128,ASUC-0129,ASUC-0130,ASUC-0131, ASUC-0132,ASUC-0133,ASUC-0134,ASUC-0135,ASUC-0136,ASUC-0137,ASUC-0138,ASUC-0139, ASUC-0140,ASUC-0141,ASUC-0142,ASUC-0143,ASUC-0143,ASUC-0144,ASUC-0145,ASUC-0162, ASUC-0163,ASUC-0164,ASUC-0165,ASUC-0166,ASUC-0167,ASUC-0168,ASUC-0169,ASUC-0170, ASUC-0171,ASUC-0172,ASUC-0173,ASUC-0203,ASUC-0204,ASUC-0205,ASUC-0206,ASUC-0207, ASUC-0208,ASUC-0209,ASUC-0210,ASUC-0218,ASUC-0219,ASUC-0220,ASUC-0221,ASUC-0222, ASUC-0223,ASUC-0224,ASUC-0225,ASUC-0235,ASUC-0236,ASUC-0237,ASUC-0340,ASUC-0341 ASUD-0244,ASUD-0261,ASUD-0262,ASUD-0263,ASUD-0267,ASUD-0281,ASUD-0282,ASUD-0283  Serial Numbers Distributed OUS ASUA-0353,ASUA-0441,ASUA-0447,ASUA-0472,ASUA-0478,ASUA-0496,,SUA-0497,ASUA-0500, ASUB-0004,ASUB-0024,ASUB-0025,ASUB-0030,ASUB-0033,ASUB-0034,ASUB-0035,ASUB-0038, ASUB-0039,ASUB-0040,ASUB-0041,ASUB-0043,ASUB-0044,ASUB-0046,ASUB-0048,ASUB-0063, ASUB-0064,ASUB-0066,ASUB-0067,ASUB-0069,ASUB-0073,ASUB-0074,ASUB-0075,ASUB-0076, ASUB-0077,ASUB-0095,ASUB-0096,ASUB-0097,ASUB-0098,ASUB-0099,ASUB-0100,ASUB-0101, ASUB-0102,ASUB-0104,ASUB-0105,ASUB-0107,ASUB-0108,ASUB-0109,ASUB-0112,ASUB-0113, ASUB-0114,ASUB-0115,ASUB-0116,ASUB-0118,ASUB-0120,ASUB-0121,ASUB-0122,ASUB-0123, ASUB-0124,ASUB-0128,ASUB-0129,ASUB-0130,ASUB-0136,ASUB-0137,ASUB-0139,ASUB-0140, ASUB-0141,ASUB-0143,ASUB-0144,ASUB-0145,ASUB-0146,ASUB-0147,ASUB-0148,ASUB-0149, ASUB-0150,ASUB-0151,ASUB-0152,ASUB-0153,ASUB-0154,ASUB-0155,ASUB-0156,ASUB-0157, ASUB-0158,ASUB-0159,ASUB-0160,ASUB-0161,ASUB-0162,ASUB-0163,ASUB-0177,ASUB-0178, ASUB-0188,ASUB-0189,ASUB-0190,ASUB-0191,ASUB-0192,ASUB-0193,ASUB-0194,ASUB-0195, ASUB-0196,ASUB-0197,ASUB-0198,ASUB-0199,ASUB-0201,ASUB-0202,ASUB-0203,ASUB-0204, ASUB-0205,ASUB-0206,ASUB-0207,ASUB-0208,ASUB-0209,ASUB-0210,ASUB-0211,ASUB-0223, ASUB-0224,ASUB-0225,ASUB-0226,ASUB-0227,ASUB-0229,ASUB-0230,ASUB-0231,ASUB-0232, ASUB-0233,ASUB-0234,ASUB-0235,ASUB-0236,ASUB-0237,ASUB-0238,ASUB-0240,ASUB-0243, ASUB-0245,ASUB-0252,ASUB-0253,ASUB-0254,ASUB-0255,ASUB-0256,ASUB-0257,ASUB-0258, ASUB-0260,ASUB-0261,ASUB-0262,ASUB-0263,ASUB-0264,ASUB-0265,ASUB-0271,ASUB-0272, ASUB-0273,ASUB-0276,ASUB-0278,ASUC-0001ASUC-0004,ASUC-0016,ASUC-0017,ASUC-0018, ASUC-0019,ASUC-0020,ASUC-0022,ASUC-0024,ASUC-0026,,ASUC-0028,,ASUC-0030,ASUC-0031, ASUC-0032,ASUC-0033,ASUC-0034,ASUC-0035,ASUC-0036,ASUC-0039,ASUC-0061,ASUC-0062, ASUC-0063,ASUC-0226,ASUC-0227,ASUC-0228,ASUC-0229,ASUC-0230,ASUC-0232,ASUC-0233, ASUC-0234,ASUC-0247,ASUC-0248,ASUC-0249,ASUC-0250,ASUC-0251,ASUC-0252,ASUC-0253, ASUC-0254,ASUC-0255,ASUC-0256,ASUC-0257,ASUC-0262,ASUC-0263,ASUC-0264,ASUC-0265, ASUC-0266,ASUC-0267,ASUC-0269,ASUC-0272,ASUC-0273,ASUC-0274,ASUC-0275,ASUC-0276, ASUC-0277,ASUC-0278,ASUC-0279,ASUC-0284,ASUC-0286,ASUC-0287,ASUC-0288,ASUC-0289, ASUC-0290,ASUC-0291,ASUC-0292,ASUC-0293,ASUC-0294,ASUC-0295,ASUC-0296,ASUC-0300, ASUC-0301,ASUC-0302,ASUC-0303,ASUC-0304,ASUC-0305,ASUC-0306ASUC-0307,ASUC-0308, ASUC-0309,ASUC-0310,ASUC-0311,ASUC-0314,ASUC-0315,ASUC-0316,ASUC-0319,ASUC-0320, ASUC-0321,ASUC-0322,ASUC-0323,ASUC-0339,ASUC-0342,ASUC-0343,ASUC-0344,ASUC-0345, ASUC-0346,ASUC-0348,ASUC-0349,ASUC-0350,ASUC-0351,ASUC-0352,ASUC-0353,ASUC-0354, ASUC-0355,ASUC-0356,ASUC-0357,ASUC-0358,ASUC-0359,ASUC-0360,ASUC-0361,ASUC-0362, ASUC-0363,ASUC-0364,ASUC-0365,ASUC-0366,ASUC-0367,ASUC-0380,ASUC-0381,ASUC-0382, ASUC-0383,ASUC-0384,ASUC-0385,ASUC-0386,ASUC-0387,ASUC-0391,ASUC-0392,ASUC-0393, ASUC-0394,ASUC-0395,ASUC-0396,ASUC-0397,ASUC-0398,ASUC-0399,ASUC-0400,ASUC-0401, ASUC-0402,ASUC-0403,ASUC-0404,ASUC-0405,ASUC-0406,ASUC-0407,ASUC-0408,ASUC-0409, ASUC-0410,ASUC-0416,ASUD-0001,ASUD-0002,ASUD-0004,ASUD-0005,ASUD-0006,ASUD-0007, ASUD-0008,ASUD-0009,ASUD-0010,ASUD-0011,ASUD-0012,ASUD-0013,ASUD-0014,ASUD-0015, ASUD-0016,ASUD-0018,ASUD-0019,ASUD-0020,ASUD-0022,ASUD-0023,ASUD-0041,ASUD-0042, ASUD-0043,ASUD-0044,ASUD-0046,ASUD-0047,ASUD-0048,ASUD-0049,ASUD-0050,ASUD-0051, ASUD-0057,ASUD-0059,ASUD-0060,ASUD-0061,ASUD-0062,ASUD-0063,ASUD-0064,ASUD-0065, ASUD-0066,ASUD-0074,ASUD-0075,ASUD-0077,ASUD-0078,ASUD-0079,ASUD-0080,ASUD-0081, ASUD-0098,ASUD-0099,ASUD-0100,ASUD-0101,ASUD-0102,ASUD-0103,ASUD-0104,ASUD-0105, ASUD-0106,ASUD-0107,ASUD-0108,ASUD-0109,ASUD-0126,ASUD-0127,ASUD-0128,ASUD-0129, ASUD-0130,ASUD-0131,ASUD-0133,ASUD-0135,ASUD-0136,ASUD-0140,ASUD-0141,ASUD-0150, ASUD-0168,ASUD-0173,ASUD-0174,ASUD-0175,ASUD-0181,ASUD-0182,ASUD-0183,ASUD-0184, ASUD-0185,ASUD-0186,ASUD-0196,ASUD-0198,ASUD-0199,ASUD-0200,ASUD-0201,ASUD-0202, ASUD-0203,ASUD-0204,ASUD-0205,ASUD-0206,ASUD-0219,ASUD-0220,ASUD-0221,ASUD-0222, ASUD-0223,ASUD-0224,ASUD-0225,ASUD-0226,ASUD-0227,ASUD-0228,ASUD-0238,ASUD-0239, ASUD-0245,ASUD-0246,ASUD-0247,ASUD-0248,ASUD-0249,ASUD-0250,ASUD-0251,ASUD-0252, ASUD-0253,ASUD-0258,ASUD-0259,ASUD-0260,ASUD-0273,ASUD-0274

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Devices were imported from Draegerwerk Ag and Co. KGaA and distributed in the following states: AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, KY, MA, MN, MO, NC, NE, NM, NV, NY, OH, PA, TX, UT, VA, WI. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Belgium, Brazil, Chili, China, Croatia, Denmark, Ecuador, France, Georgia, Germany, Greece, Hungary, India, Indonesia, Italy, Japan, Jordan, Kuwait, Lithuania, Mexico, Netherlands, Peru, Poland, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam.

Recall record

Recalling firm
Draeger, Inc.
Product type
Devices
Recall number
Z-2616-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Telford, PA
Recall initiated
2025-09-09

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2616-2025 · Source: openFDA enforcement.