  "@context": {
    "schema": "https://schema.org/",
    "au": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native#",
    "@vocab": "https://schema.org/",
    "claims": {
      "@id": "au:claims",
      "@container": "@list"
    },
    "epistemicStatus": "au:epistemicStatus",
    "occurred": "au:occurred",
    "supported": "au:supported",
    "adjudged": "au:adjudged",
    "operative": "au:operative",
    "grounding": "au:grounding",
    "groundingType": "au:groundingType",
    "adjudicator": "au:adjudicator",
    "source": "au:source",
    "sourceUrl": "au:sourceUrl",
    "value": "au:value",
    "claimType": "au:claimType",
    "reviewStatus": "au:reviewStatus",
    "commercialInfluence": "au:commercialInfluence",
    "falsifier": "au:falsifier"
  },
  "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-2638-2025.jsonld",
  "@type": "schema:Dataset",
  "name": "Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).",
  "description": "FDA recall Z-2638-2025 — Bard Peripheral Vascular Inc is recalling Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively). (Devices, Class II). The enforcement record is RECORDED; the classification is a DETERMINATION (FDA's regulatory verdict); the boundary is INTERPRETATION (a recall ≠ harm to a specific person). Source: openFDA. CC0.",
  "license": "https://creativecommons.org/publicdomain/zero/1.0/",
  "isAccessibleForFree": true,
  "isPartOf": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
  "rdfs:isDefinedBy": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native",
  "seeAlso": "https://agenticu.ryanjhunter.workers.dev/commons/the-research-object",
  "dateModified": "2026-06-24T07:02:03.626072+00:00",
  "isBasedOn": {
    "@type": "schema:Dataset",
    "name": "openFDA drug & device enforcement (recall) reports",
    "url": "https://open.fda.gov/apis/drug/enforcement/",
    "creator": {
      "@type": "schema:GovernmentOrganization",
      "name": "U.S. Food & Drug Administration"
    }
  },
  "sameAs": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-2638-2025&lang=eng",
  "about": {
    "@type": "schema:MedicalDevice",
    "name": "Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).",
    "manufacturer": {
      "@type": "schema:Organization",
      "name": "Bard Peripheral Vascular Inc"
    }
  },
  "claims": [
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-2638-2025/enforcement-record",
      "@type": "schema:Claim",
      "name": "FDA recall Z-2638-2025: Bard Peripheral Vascular Inc is recalling Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).",
      "text": "The FDA logged enforcement record Z-2638-2025: Bard Peripheral Vascular Inc is voluntarily recalling Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively). — reason: Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a  Red X  indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.. Status: Ongoing — the recall is still being carried out.. Initiated 2025-08-21. Affected lots/codes: UDI-DI: 00858254006022. Generator Software Version 3.35. A RECORDED FDA enforcement action (occurred=true).",
      "claimType": "official_fact",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": true,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "SourceCitation",
        "source": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-2638-2025&lang=eng",
        "value": "Devices recall, status Ongoing",
        "adjudicator": {
          "name": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-2638-2025&lang=eng"
        }
      },
      "falsifier": "A RECORD of measured/reported events (occurred=true) from openFDA drug & device enforcement (recall) reports — U.S. FDA; superseded when openFDA drug & device enforcement (recall) reports — U.S. FDA publishes its next release."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-2638-2025/fda-classification",
      "@type": "schema:Claim",
      "name": "FDA classified recall Z-2638-2025 as Class II",
      "text": "The FDA assigned this recall Class II: Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.. This is the FDA's official risk classification — its regulatory DETERMINATION of the hazard category (a verdict constituted by the FDA's recognition, occurred=FALSE), adjudged=TRUE because the FDA issued it. It is NOT a measurement of harm to any specific person.",
      "claimType": "regulatory_determination",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "AdjudicativeAct",
        "source": "U.S. FDA recall classification (openFDA enforcement)",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-2638-2025&lang=eng",
        "value": "Class II",
        "adjudicator": {
          "name": "U.S. FDA recall classification (openFDA enforcement)",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-2638-2025&lang=eng"
        }
      },
      "falsifier": "occurred=FALSE: a regulatory DETERMINATION (an authority's classification/judgment) has NO material truthmaker independent of recognition — it is constituted by U.S. FDA recall classification (openFDA enforcement)'s say-so (a Searle institutional fact), not measured. adjudged=TRUE because U.S. FDA recall classification (openFDA enforcement) (an external authority) issued it. Distinct from a recorded measurement (which has an independent truthmaker) and from a modeled figure (a computed quantity, not a verdict). Challengeable; superseded if the authority reclassifies."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-2638-2025/boundary",
      "@type": "schema:Claim",
      "name": "What an FDA recall is — and is NOT",
      "text": "A recall is a RECORDED FDA enforcement action — the FDA logged that a firm is removing or correcting a product, classed by the FDA's risk judgment. It does NOT establish that any specific unit is affected or caused harm; causation is not established. This record informs and compares — it is NOT medical advice and does not tell anyone to start or stop a drug or device. For a medical decision, route to the FDA recall record and a licensed pharmacist or prescriber.",
      "claimType": "interpretation",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": false,
        "operative": true
      },
      "grounding": {
        "groundingType": "Interpretation",
        "source": "Atlas Commons L5 boundary (enforcement-action ≠ harm causation)",
        "sourceUrl": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
        "value": "a recorded enforcement action, not proof of harm or causation; not medical advice",
        "adjudicator": null
      },
      "falsifier": "occurred=FALSE (a framing, not an event) and adjudged=FALSE (an in-house reading, not an authority's verdict) — the LOWEST-warrant claim here. A flat representation would render this identically to the official facts; the four-status does not."
    }
  ],
  "hasPart": [
    {
      "@type": "schema:ItemList",
      "name": "FDA recall details — Z-2638-2025",
      "numberOfItems": 9,
      "itemListElement": [
        {
          "@type": "schema:ListItem",
          "position": 1,
          "item": {
            "Field": "Recall number",
            "Value": "Z-2638-2025",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 2,
          "item": {
            "Field": "Product type",
            "Value": "Devices",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 3,
          "item": {
            "Field": "Classification",
            "Value": "Class II",
            "Status": "determination"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 4,
          "item": {
            "Field": "Status",
            "Value": "Ongoing — the recall is still being carried out.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 5,
          "item": {
            "Field": "Recalling firm",
            "Value": "Bard Peripheral Vascular Inc",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 6,
          "item": {
            "Field": "Reason",
            "Value": "Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a  Red X  indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 7,
          "item": {
            "Field": "Initiated",
            "Value": "2025-08-21",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 8,
          "item": {
            "Field": "Lot/code info",
            "Value": "UDI-DI: 00858254006022. Generator Software Version 3.35",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 9,
          "item": {
            "Field": "Distribution pattern",
            "Value": "Worldwide - US Nationwide distribution including in the states of NJ, NY, FL, LA, VA, CA, TX, NV, PA, NE, TN, IL, AZ, SC, AK, GA, CO, HI, OH, NC, WI, WV, MD, MO, AL, OK, MS, PR, KS, UT, CT, IA, MI, IN, MA, KY, WA, MT, MN, SD, ME, DE, NM, AR and the country of Malaysia.",
            "Status": "recorded"
          }
        }
      ]
    }
  ]
}