Class II FDA device recall · Z-2648-2023

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

(1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica MAGNETOM Spectra Siemens Model Number:11060845 Used with the following MAGNETOM MRI Systems: MAGENTOM Amira 10836777 & 10838610; MAGENTOM Free.Max 11408424 & 11408425; MAGNETOM Free.Star 11408426 & 11408427; MAGNETOM Mica 11292557; MAGNETOM Sempra 10840815 & 10840816; MAGNETOM Spectra 10655588 & 10837643; (2) Over Ear Headphones SMN: 11060845 used with the following MAGNETOM MRI systems: MAGENTOM Aera 10432914 MAGENTOM Altea 13344915 & 11410371 MAGNETOM Avanto 10849579 MAGNETOM Avanto Fit 11516216 MAGNETOM AvantoFit 10849578 MAGNETOM Cima.X 11647156 MAGNETOM Lumina 11344916 & 11516153 MAGNETOM Prisma 10849582 MAGNETOM PrismaFit 10849583 MAGNETOM Skyra 10432915 & 10849580 MAGNETOM Skrya Fit 11516217 MAGNETOM Sola 11291455 & 11410231 MAGNETOM Sola Fit 11410482 MAGNETOM Terra 10882764 MAGNETOM Terra.X 11371477 MAGNETOM Vida 11060815 & 11516152 MAGNETOM VidaFit 11410481

Why it was recalled

In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential artifacts depend on the position of the headband, the amount of lubricating grease, as well as the sequence parameters used

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
(1)Model/Material Number/UDI: MAGNETOM FREE.MAX 11408425   UDI: 04056869281742; MAGNETOM AMIRA 10836777  UDI: 04056869006819; MAGENTOM AMIRA 10838610 UDI: 04056869020488; MAGNETOM FREE.Max 11408424 UDI: 04056869281759; MAGNETOM Free.Star 11408426  UDI: 04056869281766; MAGNETOM Free.Star 11408427 UDI:  0405686981773; MAGNETOM Mica 11292557 NA; MAGNETOM Sempra 10840815  UDI: 04056869044521.  Over Ear Headphones   11060845 (Serial Numbers 7981 to 32200Serial Numbers; 7981 to 32200  (2) Model UDI-DI MAGNETOM AERA 10432914 04056869006697 MAGNETOM ALTEA 11344915 04056869230719  11410371 04056869230726 MAGNETOM AVANTO FIT 11516216 04056869299266 MAGNETOM AVANTOFit   10849578 04056869006642  10849579 04056869006659 MAGENTOM CIMA.X 11647158 04056869975221¿  11689304 04056869975245 MAGNETOM LUMINA 11344916 04056869230740  11516153 04056869260136 MAGENTOM PRISMA 10849582 04056869006727 MAGENTOM PRISMAFit 10849583 04056869006734 MAGNETOM SKYRA 10432915 04056869006703 MAGENTOM SKYRA 10849580 04056869006710 MAGENTOM SKYRA FIT 11516217 04056869299273 MAGNETOM SOLA 11291455 04056869164809  11410231 04056869217864 MAGNETOM SOLA FIT 11410482 04056869245195 MAGNETOM TERRA 10882764 04056869039190 MAGNETOM TERRA.X 11371477 04056869039190 MAGNETOM Viato.Mobile 11689652 04056869985558 MAGNETOM VIDA 11060815 04056869039176                           11516152 04056869260143 MAGNETOM VIDAFit 11410481 04056869245188   Over Ear Headphones   11060845 (Serial Numbers 7981 to 32200

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide distribution.

Recall record

Recalling firm
Siemens Medical Solutions USA, Inc
Product type
Devices
Recall number
Z-2648-2023
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Origin
Malvern, PA
Recall initiated
2023-09-01

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-2648-2023 · Source: openFDA enforcement.