Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter Instructions for Use for directions on the use of these accessories.
Why it was recalled
Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur if the user attempts to insert the proximal end into the patient.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
Catalog Number: 5583705 Lot Number / UDI codes: REGZ2564 (01)00801741085451(17)271130(10)REGZ2564 REFV0870 (01)00801741085451(17)260731(10)REFV0870 REFU4050 (01)00801741085451(17)260630(10)REFU4050 REFT3309 (01)00801741085451(17)260531(10)REFT3309 REEX1002 (01)00801741085451(17)250930(10)REEX1002 REEW2740 (01)00801741085451(17)250831(10)REEW2740 REEV0897 (01)00801741085451(17)250731(10)REEV0897 REDY2576 (01)00801741085451(17)241031(10)REDY2576 REEU1584 (01)00801741085451(17)250630(10)REEU1584 REER0990 (01)00801741085451(17)250331(10)REER0990 REEQ0825 (01)00801741085451(17)250228(10)REEQ0825
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide - U.S. Nationwide distribution in the states of OH, TX, and WV. The countries of Belgium, Canada, France, Italy, and United Kingdom.
Recall record
- Recalling firm
- Bard Peripheral Vascular Inc
- Product type
- Devices
- Recall number
- Z-3062-2024
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Tempe, AZ
- Recall initiated
- 2024-07-10
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·