Drug & Device Safety Atlas
Class II FDA device recall · Z-3249-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)

Why it was recalled

During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
IntelliVue Patient Monitor MX400, 866060, UDI-DI: 00884838038752; IntelliVue Patient Monitor MX430, 866061, UDI-DI: 00884838057562; IntelliVue Patient Monitor MX450, 866062, UDI-DI: 00884838038769; IntelliVue Patient Monitor MX500, 866064, UDI-DI: 00884838038776; IntelliVue Patient Monitor MX550, 866066, UDI-DI: 00884838038783;  Only MX400-550 devices shipped after 26-April-2024 are affected. Please refer to the manufacturing date on the back of your monitor.

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide. Global Distribution.

Recall record

Recalling firm
Philips North America
Product type
Devices
Recall number
Z-3249-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Cambridge, MA
Recall initiated
2024-08-30

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-3249-2024 · Source: openFDA enforcement.