Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Proboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry System (Product Code 6801375), VITROS 5,1 Chemistry System (Refurbished) (Product Code 6801890), VITROS 3600 Immunodiagnostic System (Product Code 6802783), VITROS 3600 Immunodiagnostic System (Refurbished) (Product Code 6802914), VITROS 4600 Chemistry System (Product Code 6802445), VITROS 4600 Chemistry System (Refurbished) (Product Code 6900440), VITROS 5600 Integrated System (Product Code 6802413), VITROS 5600 Integrated System (Refurbished) (Product Code 6802915), VITROS XT 3400 Chemistry System (Product Code 6844458), and VITROS XT 7600 Integrated System (Product Code 6844461)
Why it was recalled
Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an out-of-specification barrel component. This may cause VITROS VersaTips and/or VITROS MicroTips to be improperly seated on the proboscis, which may result in falsely elevated Gentamicin and Valproic Acid assay results as well as an increased frequency of condition codes resulting in delayed patient results. The issue was identified by internal testing when reported instruments failed to pass release testing and were traced to the molded Proboscis part J56639 with date codes from 040722 onward.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
Assemblies with a date code between 040722 - 121922
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide - US Nationwide distribution.
Recall record
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product type
- Devices
- Recall number
- Z-0008-2024
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Rochester, NY
- Recall initiated
- 2023-08-09
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.