Drug & Device Safety Atlas
Class II FDA device recall · Z-0638-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2110X11B EE2100X15B EE2110X15B ES2100X15B ES2110X15B EU2100X15B EU2100X19 EU2110X15B FR2100X14B FR2110X14B FX2100X15B GB2110X15B IA2100X15B IA2110X15B IN2100X15B IN2100X19 IN2110X15B IT2100X21B IT2110X21B JP2100X16B JP2110X16B KR2110X15B LA2100X15B LA2110X15B ND2110X15B RDE2100X13B RDE2110X13B RDS2110X11B RIN2100X15B RIN2110X15B RJP2100X16B RJP2110X16B UDS2110X11B

Why it was recalled

The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Material   UDI-DI/GTIN BL2110X15B 606959055520 BR2110X18B 606959055476 CA2110X12B 606959052307 DE2100X13B 606959054103 DE2110X13B 606959054110 DS2100X11B 606959051997 DS2110X11B 606959051942 EE2100X15B 606959055698 EE2110X15B 606959055544 ES2100X15B 606959054097 ES2110X15B 606959054080 EU2100X15B 606959055674 EU2100X19 606959055681 EU2110X15B 606959054981 FR2100X14B 606959055643 FR2110X14B 606959055506 FX2100X15B 606959061019 GB2110X15B 606959054127 IA2100X15B 606959055636 IA2110X15B 606959055490 IN2100X15B 606959054059 IN2100X19 606959055575 IN2110X15B 606959051959 IT2100X21B 606959055438 IT2110X21B 606959055421 JP2100X16B 606959055582 JP2110X16B 606959051973 KR2110X15B 606959055483 LA2100X15B 606959055599 LA2110X15B 606959055452 ND2110X15B 606959055513 RDE2100X13B  RDE2110X13B  RDS2110X11B 606959060517 RIN2100X15B  RIN2110X15B  RJP2100X16B  RJP2110X16B  UDS2110X11B 606959060500  All units of listed models are affected by this correction.

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide distribution: US (nationwide)-AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Including PR; and OUS (International) including countries of: AE, AR,AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, ;EC EE, EG, ES, FI, FR, GB, GE, GF, GR, GT, HK, HR, HU, IE, IL, IN, IT, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, YT, ZA and ZW.

Recall record

Recalling firm
Philips Respironics, Inc.
Product type
Devices
Recall number
Z-0638-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Murrysville, PA
Recall initiated
2023-11-27

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0638-2024 · Source: openFDA enforcement.