Drug & Device Safety Atlas
Class II FDA device recall · Z-0825-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; c) MINI MINOR AGH ASC PACK, kit number AGMM60M; d) OUTPATIENT PORT ACCESS KIT 307907, kit number AHPO55J; e) General Pack, kit number BBGP31C; f) FRACTURE TABLE PACK, kit number FDFT75N; g) Arthroscopy Pack, kit number HCAR04K; h) BARIATRIC PACK, kit number HGBA92F; i) LAPAROTOMY PACK, kit number HNLP12J; j) MIDLINE CHANGE KIT, kit number HSMD20E; k) Minor General, kit number JSMI10I; l) ENDO VASCULAR PACK, kit number LVEV61; m) MAJOR PACK, kit number LVMJ21; n) HEAD AND NECK PACK, kit number MMHN37F; o) KIT, MINOR U BAR, kit number MMNU34E; p) KIT, MAJOR UNIVERSAL, kit number MMUN28E; q) VAG PLASTY PACK, kit number OSVP16; r) DR L MAJOR, kit number SALM40H; s) DR L MAJOR, kit number SALM40I; t) MINOR BASIN SAH, kit number SAMB25E; u) MAJOR SAH, kit number SAMJ36G; v) MINOR PACK, kit number SMMN65; w) General Pack, kit number THGN59D; x) MINOR PACK, kit number TPMN15; y) MAJOR UNIVERSAL PACK, kit number TPUN48

Why it was recalled

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
a) AGBR90E, UDI/DI 00191072036865, Lot Numbers:  959231, exp. 02/10/2024; 681221, exp. 11/15/2023; 721221, exp. 09/30/2023; 771221, exp. 08/147/2023; 827221, exp. 06/22/2023; 897221, exp. 04/13/2023;  b) AGMM60M, UDI/DI 00191072063786, Lot Numbers:  939231, exp. 03/02/2024; 638221, exp. 4528812/28/2023; 694221, exp. 11/02/2023; 734221, exp. 09/23/2023; 769221, exp. 08/19/2023;  c) AHPO55J, UDI/DI 00191072171566, Lot Numbers:  682221, exp. 11/14/2023; 661221, exp. 12/5/2023; 977231, exp. 1/23/2024; 945231, exp. 2/24/2024; 891231, exp. 4/19/2024;  d) BBGP31C, UDI/DI none, Lot Numbers:  687221, exp. 11/9/2023; 738221, exp. 9/19/2023; 805221, exp. 7/14/2023; 870221, exp. 5/10/2023;  e) FDFT75N, UDI/DI none, Lot Numbers:  773221, exp. 8/15/2023; 714221, exp. 10/13/2023; 990231, exp. 1/10/2024; 935231, exp. 3/6/2024; 890231, exp. 4/20/2024; 870231, exp. 5/10/2024;  f) HCAR04K, UDI/DI 00191072114174, Lot Numbers:  757221, exp. 08/31/2023;  g) HGBA92F, UDI/DI none, Lot Numbers:  757221, exp. 8/31/2023; 695221, exp. 11/1/2023; 635221, exp. 12/31/2023; h) HNLP12J, UDI/DI 00191072146274, Lot Numbers:  805221, exp. 7/14/2023; 742221, exp. 9/15/2023; 673221, exp. 11/23/2023; 924231, exp. 3/17/2024; i) HSMD20E, UDI/DI 00191072130068, Lot Numbers:  898221, exp. 4/12/2023; 772221, exp. 8/16/2023; 984231, exp. 1/16/2024; j) JSMI10I, UDI/DI none, Lot Numbers:  766221, exp. 08/22/2023;   k) LVEV61, UDI/DI 00191072033475, Lot Numbers:  678221, exp. 11/18/2023; 637221, exp. 12/29/2023; l) LVMJ21, UDI/DI 00191072139023, Lot Numbers:  833221, exp. 6/16/2023; 764221, exp. 8/24/2023; 717221, exp. 10/10/2023; 988231, exp. 1/12/2024; 954231, exp. 2/15/2024; m) MMHN37F, UDI/DI none, Lot Numbers:  639221, exp. 12/27/2023;  n) MMNU34E, UDI/DI none, Lot Numbers:  928231, exp. 3/13/2024; 889231, exp. 4/21/2024; o) MMUN28E, UDI/DI none, Lot Numbers:  889231, exp. 04/21/2024;  p) OSVP16, UDI/DI 00191072156570, Lot Numbers:  827221, exp. 6/17/2023;  q) OSVP16, UDI/DI 00191072170835, Lot Numbers:  700221, exp. 10/23/2023; 644221, exp. 12/4/2023; 987231, exp. 11/20/2023; r) SALM40H, UDI/DI none, Lot Numbers:  826221, exp. 6/23/2023; 888221, exp. 4/22/2023; s) SALM40I, UDI/DI none, Lot Numbers:  687221, exp. 11/9/2023; 756221, exp. 9/1/2023; 984231, exp. 1/16/2024;  t) SAMB25E, UDI/DI none, Lot Numbers:  843221, exp. 6/6/2023; 894221, exp. 4/16/2023; 871221, exp. 5/9/2023; 686221, exp. 11/10/2023; 758221, exp. 8/30/2023; 793221, exp. 7/26/2023; 974231, exp. 1/26/2024; u) SAMJ36G, UDI/DI none, Lot Numbers:  660221, exp. 12/6/2023; 704221, exp. 10/23/2023; 752221, exp. 9/5/2023; 800221, exp. 7/19/2023; v) SMMN65, UDI/DI 00191072188472, Lot Numbers:  969231, exp. 1/31/2024; 926231, exp. 3/15/2024; w) SMMN65, UDI/DI none, Lot Numbers:  919231, exp. 3/22/2024;  x) THGN59D, UDI/DI 00191072191069, Lot Numbers:  805221, exp. 7/14/2023; 721221, exp. 10/6/2023; y) TPMN15, UDI/DI 00191072171771, Lot Numbers:  737221, exp. 9/20/2023; 702221, exp. 10/25/2023; 664221, exp. 12/2/2023; 990231, exp. 1/10/2024; 960231, exp. 2/9/2024; 939231, exp. 3/2/2024; 903231, exp. 4/7/2024;  z) TPUN48, UDI/DI 00191072171832, Lot Numbers:  738221, exp. 9/19/2023; 688221, exp. 11/8/2023; 637221, exp. 12/29/2023; 967231, exp. 2/2/2024; 924231, exp. 3/17/2024

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide distribution.

Recall record

Recalling firm
American Contract Systems, Inc.
Product type
Devices
Recall number
Z-0825-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
E-Mail
Origin
Temple Terrace, FL
Recall initiated
2023-11-29

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0825-2024 · Source: openFDA enforcement.